Device Errors 

Error or symbol 

Cause 

Resolution 

Device shows “E2”  

Cuff is under‐inflated or 
over‐inflated. 

Single Mode: 
The monitor will attempt 1 reading. 
 
TruRead™ Mode: 
The monitor will attempt up to 5 times  
 
After the automatic attempts, remove the 
arm cuff.  Read, “Taking a Measurement” 
in the instruction manual.  Take another 
measurement. 

Device shows ‘E1’ 

Air plug is not connected or 
air is leaking from the arm 
cuff. 

Check air plug connection.  Take another 
measurement and remain still until the 
measurement is complete. 

Device shows “Battery 
picture with a cross mark” 

Batteries are worn 

Replace the four batteries. 
Refer to “Battery Installation” in the 
instruction manual. 

No Power 
 
No Display appears on the 
unit 

Power off 

Check that the device is turned on.  
 
Check the battery installation for proper 
placement of the battery polarities. 
 
Replace all four batteries with new ones. 
 
If running from a wall outlet, check that 
the cord is attached to both the device 
and outlet