beautypg.com

Specifications – Omron Healthcare HEM-422CRLC User Manual

Page 18

background image

18

SPECIFICATIONS

Model:

HEM-422CRLC

Display:

LCD Digital Display

Measurement Range:

Pressure: 0 to 280 mmHg
Pulse: 40-180/min.

Accuracy/Calibration:

Pressure: ±3mmHg or 2% of reading
Pulse: ±5% of reading

Inflation:

Manual by inflation bulb

Deflation:

Automatic pressure release valve

Rapid Air Release:

Push button valve

Pressure Detection:

Capacitive pressure sensor

Measurement Method:

Oscillometric method

Pulse Wave Detection:

Capacitive pressure sensor

Power Source:

4 "AA" batteries (not included)

Battery Life:

Approximately 1000 uses

Operating Temperature/

50°F to 104°F (10°C to 40°C)

Humidity:

30 to 85% RH maximum

Storage Temperature/

-4°F to 140°F (-20°C to 60°C)

Humidity:

10 to 95% RH maximum

Console Weight:

Approximately 13 oz. (350g) with cuff,
exclude batteries

Outer Dimensions:

Approximately 4 1/8" (w) x 3 3/8" (h)
x 4 1/2" (d) (106mm x 84mm x 116mm

)

Cuff Dimensions:

Approximately 6 7/8" (w) x 24 1/2" (l)
(175mm x 622mm)

Arm Circumference Range:

Approximately 13'' to 17'' (330mm - 432mm)

Accessories:

Large Arm Cuff, Inflation Bulb,
Instruction Manual

UPC Code:

0 73796 42223 3

NOTE: These specifications, to improve performance, are subject to change without notice.