beautypg.com

Temperature ranges, Signal conditioners – Omega Vehicle Security LDP1XX-21 User Manual

Page 10

background image

Page : 9

9.- TEMPERATURE RANGES

± 0.5 %

C

B

A

PT1. ZERO adjust

COMPONENTS SIDE

PT2. SPAN adjust.

The control board contains the Analog/Digital con-
verter, the Signal Conditioner and other compo-
nents.
The function of the Signal Conditioner is amplifier
and make the linearisation of the signal received
from the sensor and there is one Signal Conditioner
for every model of RTD or Thermocouple sensor.
The trimmers for Span and Zero are also located on
the Signal Conditioner. Use these trimmers only for
calibration procedure.

Signal conditioner
Components side.

Components view

ALL MODELS
0°C to + 50°C
32°F to 122°F

Control board. Horizontal position in instruments serie LDP-124

Vertical position in instruments serie LDP-144

JG JE JF

Decimal Point
Selector

JG = XXX.X
JE = X.XXX
JF = XX.XX

10.1.- SIGNAL CONDITIONER FOR RTD SENSOR

C

B

A

PT1

ZERO adjust.

COMPONENTS SIDE

Signal

Instrument

Conditioner

Reference

Reference

MB-21

LDP-1X4-M1 *

MB-21-F2

LDP-1X4-M2

MB-21-F3

LDP-1X4-M3 *

PT2 SPAN adjust.

All jumpers on Decimal Point Selector must be opened.

10.2.- SIGNAL CONDITIONERS FOR THERMOCOUPLE SENSORS

9.1.- MODEL M1, M2 & M3. RTD SENSOR (PT-100)

TEMPERATURE ACCURACY MODEL TEMPERATURE ACCURACY RESOLUTION
RANGE

RANGE

Signal

Instrument

Conditioner Reference
Reference

MB-22 or 22F LDP-1X4-J1,J2
MB-23 or 23F LDP-1X4-K1,K2
MB-24 or 24F LDP-1X4-T1,T2
MB-25 or 25F LDP-1X4-E1,E2
MB-26 or 26F LDP-1X4-S1,S2
MB-27 or 27F LDP-1X4-R1,R2
MB-28

LDP-1X4-L1

MODEL

TEMPERATURE

ACCURACY

RESOLUTION MAXIMUM

RANGE

(±1 COUNT)

COMPENSATION

M1

- 100.0 to + 650.0 °C

0.1 °C

M2

-300 to 1526 °F

1 °F

10

M3

- 200.0 to 999.9 °F

0.1 °F

Allowable error in negative range due to the lead wire resistance : 0.2°C/

or 0.4°F/

9.2.- MODELS: J1,J2,K1,K2,T1,T2, E1,E2,S1,S2,R1,R2 & L1. THERMOCOUPLE SENSORS

10.- SIGNAL CONDITIONERS.

± 1 count

± 1 count

MODEL

J 1

K1

T1

E1

S1

R 1

L1

± 0.5 %

± 0.5 %

± 0.5 %

± 0.5 %

± 0.5 %

± 0.2 %

± 0.2 %

± 0.2 %

± 0.1 %

± 0.3 %

± 0.3 %

± 1 %

J 2

K2

T2

E2

S2

R 2

1°C / 1°F

1°C / 1°F

1°C / 1°F

1°C / 1°F

1°C / 1°F

1°C / 1°F

1°C

- 50 to + 600°C

0 to + 1250°C

+ 1000 to + 1750°C

- 50 to + 400°C

0 to + 650°C

+ 970 to + 1750°C

- 50 to + 600°C

- 40 to 1000°F

32 to 2250°F

- 40 to 750°F

32 to 1200°F

1775 to 3150°F

1850 to 3150°F

AMBIENT TEMP.
COMPENSATION

± 0.2 %

± 0.4 %

± 0.4 %

* Jumper JG located on Decimal Point Selector must be closed.