Troubleshooting – Omega FPDM1000 User Manual

Page 7

background image

5

Symptom

Probable Cause

Corrective Action

FLUID WILL NOT

1. Foreign matter blocking

Dismantle meter, clean rotors. Strainer must

FLOW THROUGH

rotors

be fitted in-line.

THE METER

2. Line strainer blocked

Clean strainer.

3. Damaged rotors

Replace rotors. Strainer must be fitted in-line.

4. Meter connections over-

Re-adjust connections.

tightened

5. Fluid is too viscous

See specifications for maximum viscosity.

REDUCED FLOW

1. Strainer is partially

Clean strainer.

THROUGH THE

blocked

METER

2. Fluid is too viscous

See specifications for maximum viscosity.

METER READING

1. Fluid flowrate is too high

See specifications for minimum and maximum

INACCURATE

or too low

flowrates.

2. Fluid is too viscous

Bleed air from system.

3. Excess wear caused by

Check meter body and rotors. Replace as

incorrect installation

required. Refer to installation instructions.

FLUID FLOWS

1. Bevel gear is loose on

Tighten grub screws.

BUT NO READING

on shaft

ON METER

2. Rotor drive gear is

Replace rotor.

damaged

3. Transmission gears

Replace gears.

damaged

4. Register gears damaged

Replace register assembly.

FLUID LEAKS

1. Seal worn or damaged

Replace seal. Check seal compatibility with

INTO REGISTER

on the coverplate

fluid.

TROUBLESHOOTING