beautypg.com

ResMed VPAP S User Manual

Page 2

background image

Global leaders in sleep and respiratory medicine www.resmed.com

TECHNICAL SPECIFICATIONS

P

erformance

Operating pressure range: 3 to 25 cm H

2

O

Maximum single fault pressure: 40 cm H

2

O

Dynamic pressure characteristics
S mode:
IPAP: 4 to 25 cm H

2

O; EPAP: 3 to 25 cm H

2

O

CPAP mode: 4 to 20 cm H

2

O

Sound pressure level
< 26 dB (tested in accordance with the requirements of ISO 17510-1:2002)
Data storage
Data includes usage, leak, pressure, AHI, AI, tidal volume, respiratory rate and

minute ventilation

Via LCD menu: 365 sessions of summary data
Via ResLink: 365 sessions of summary data, 30 sessions of detailed data
Via ResScan Data Card: 180 sessions of summary data, two sessions of

detailed data

Dimensions (L x W x H): 4.4” x 6.5” x 5.7” (112 mm x 164 mm x 145 mm)

Weight: 2.9 lb (1.3 kg)

Power supply
Input range 100–240VAC, 50–60Hz, 40VA (typical power consumption),

< 100VA (maximum power consumption), 12–24DVC via ResMed DC
24V/50W converter

Environmental conditions
Operating temperature: 41°F to 97°F (+5°C to +35°C)
Operating humidity: 10–95% non-condensing

Electromagnetic compatibility
Product complies with all applicable electromagnetic compatibility

requirements (EMC) for residential, commercial and light industry
environments according to IEC 60601-1-2.

Air filter
Two-layered, powder-bonded, polyester non-woven fiber

IEC 60601-1 classification
Class II (double insulated), Type CF

Note: The manufacturer reserves the right to change these specifi cations
without notice.

PRODUCT CODES

VPAP S

26119

H4i with standard water chamber

26940

VPAP S + H4i

26120

S8 ResLink

22206

Oximeter Kit

1430050

S8 ResTraxx US

22204

S8 ResTraxx Canada

22221

ResMed Corp Poway, CA, USA +1 858 746 2400 or 1 800 424 0737 (toll free), ResMed Ltd Bella Vista, NSW, Australia +61 (2) 8884 1000 or 1 800 658 189 (toll free). Offices in Austria, Belgium, Brazil, China, Finland,
France, Germany, Greece, Hong Kong, India, Ireland, Italy, Japan, Malaysia, Mexico, Netherlands, New Zealand, Norway, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, United Kingdom (see website
for details). EPR, HumidAire, ResLink, S8, ResScan, ResTraxx, TiControl, and Vsync are trademarks of ResMed Ltd. H4i and VPAP are trademarks of ResMed Ltd and are registered in U.S. Patent and Trademark Office.
©2008 ResMed. 1011085/1 08 10

VPAP S FEATURES SUMMARY

Modes of operation

• S (Spontaneous)

• CPAP (with EPR)

Product features

• Simplified standard setting menu

• Vsync automatic leak management

• TiControl (Ti max/min)

• Adjustable breath trigger/cycle

• 24V DC input (via converter)

• Compact S8

design

• Easy-Breathe motor

Heated humidification

• Optional, fully integrated

Data storage

• Usage and treatment summary (day, week, month, six-month, year)

• Usage profile

• Summary data storage

• Detailed data storage

• Remote settings adjustment

Data transfer

• Direct download to PC

• ResLink via SmartMedia card

• ResScan Data Card

• Live oximetry with ResLink

• Remote compliance monitoring via ResTraxx