beautypg.com

Technical specifications, Pressure (cm h – ResMed Mirage Activa LT User Manual

Page 9

background image

8

Technical Specifications

The mask system does not contain latex, PVC or DEHP materials.

Pressure–Flow

Curve

The mask contains passive venting to protect against rebreathing.

As a result of manufacturing variations, the vent flow rate may

vary.

0

5

10

15

20

25

30

35

40

45

50

4

6

8

10

12

14

16

18

20

Pressure

(cm H

2

O)

Flow

(L/min)

4

19

8

28

12

34

16

40

20

45

Vent Flo

w R

ate (L/min)

Mask Pressure (cm H

2

O)

Dead Space

Information

The dead space of the mask varies according to cushion size. It is

185 mL for the Medium size.

Therapy Pressure 4 to 20 cm H

2

O

Environmental

Conditions

Operating temperature: +5°C to +40°C

Operating humidity: up to 95% non-condensing

Storage and transport temperature: -20°C to +60°C

Storage and transport humidity: up to 95% non-condensing

Sound

DECLARED DUAL-NUMBER NOISE EMISSION VALUES in ac-

cordance with ISO 4871. The A-weighted sound power level of the

mask is 30 dBA, with uncertainty 3 dBA. The A-weighted sound

pressure level of the mask at a distance of 1 m is 22 dBA, with

uncertainty 3 dBA.

Resistance

Drop in pressure measured (nominal)

at 50 L/min: 0.1 cm H

2

O

at 100 L/min: 0.7 cm H

2

O

Gross dimen-

sions

M, L, LW: 154 mm (H) x 95 mm (W) x 117 mm (D)

Mask fully assembled – no headgear.

The manufacturer reserves the right to change these specifications without

notice.