15
T
ECHNICAL
S
PECIFICATIONS
E
NG
LI
SH
T
ECHNICAL
S
PECIFICATIONS
P
ERFORMANCE
Maximum heater plate temperature: 85°C
Temperature cut-out: 93°C
Maximum output humidity: 95% relative humidity (control dial set to maximum setting,
flow rate 50 L/min, ambient temperature of 20°C, ambient relative humidity of 40%).
Maximum gas temperature: 41°C
D
IMENSIONS
(H
X
W
X
D)
Docking station and water chamber: 118 mm x 170 mm x 212 mm
W
EIGHT
Docking station and unfilled water chamber: 670 g
Water capacity to maximum fill line: 385 mL
P
OWER
S
UPPLY
Input range for R
ES
M
ED
S8 flow generator with
H
UMID
A
IRE
3i: 100–240V, 50/60Hz;
110V, 400Hz; 2.5A <140 VA (110 W) (maximum power consumption). Instantaneous
peak power consumption <340 VA.
Actual average power consumption will vary, depending on factors such as
temperature setting, set pressure, the use of accessories, height above sea level and
ambient temperature.
Maximum heater element power: 85 W.
H
OUSING
C
ONSTRUCTION
Docking station: Flame retardant engineering thermoplastic, plated cast aluminium.
Water chamber: Injection molded plastic, stainless steel.
E
NVIRONMENTAL
CONDITIONS
Operating temperature: +5°C to +40°C
Operating humidity: 10–95% non-condensing
Storage and transport temperature: -20°C to +60°C
Storage and transport humidity: 10–95% non-condensing
IEC 60601-1 C
LASSIFICATION
Class II (double insulation)
Type CF
E
LECTROMAGNETIC
C
OMPATIBILITY
Please refer to your flow generator user's manual for details.
Note: The manufacturer reserves the right to change these specifications without notice.