Clinical Use Only

4

Option 3: STERRAD Sterilization System

Note: The Ultra Mirage™ II Nasal Mask is not validated for the STERRAD
Sterilization System.

Pre-sterilization

1 Disassemble the mask.
2 Clean the individual parts of the mask with a soft bristle brush for one

minute while soaking them in detergent (eg, Alconox). Pay particular
attention to all crevices and cavities.

3 Rinse the components twice by agitating them vigorously in drinking quality

water and allow them to air dry out of direct sunlight.

ResMed has tested the detergent Alconox (diluted at 1%).

Sterilization

Sterilize the sterilizable mask components as described in the manufacturer’s
instructions for the STERRAD Sterilization System (eg, Sterrad 100S or
Sterrad NX).

Testing by ResMed has shown that the mask components can withstand 15
cycles of the STERRAD 100S Sterilization System, when using the tray
method.
Note: The use of pouches is not recommended.

Inspection after Disinfection/Sterilization

Perform a visual inspection of each mask component. If any visible
deterioration of a mask component is apparent (cracking, crazing, tears etc),
the mask component should be discarded and replaced. Slight discoloration of
the silicone components may occur and is acceptable.
Note: If a healthcare facility requires an additional disinfection or sterilization
process after reassembly, the number of validated cycles is halved.

Parts not Validated for Multipatient Use

Some parts of the masks have not been validated for multipatient use. These
parts need to be replaced with new parts when using between patients.
Please check the Components Card on the website www.resmed.com for a list
of available replacement parts for each mask system.