12-Bit Simultaneous Grayscale Display

Many current monitors use anti-glare panels, which have a waffled
surface to diffuse the reflection of surrounding light. However, this
also diffuses the light from the monitor’s backlight, which affects the
outline of an image, causing a viewing strain on the observer. With a
distinctive glare panel, there is no diffused reflection by the waffled
surface, and the outline of the image is displayed more clearly.
Therefore, the subtle images such as calcification and masses will be
easier to observe reducing the strain on the observer.

Sharper Resolution for Clearer Image Outline

MTF (Modulation Transfer Function) is a parameter to
define the display sharpness and resolution of a monitor
panel. The glare panel has better MTF characteristics than

the anti-glare panel meaning the
glare panel can be expected to
display the outline of a detailed
image more sharply.

256 Tones Image

4,096 Tones Image

12-bit graphics board and 12-bit viewer software needed for 12-bit display.

1.2

1

0.8

0.6

0.4

0.2

0

Spatial Frequency (Line pairs/mm)

Anti-Glare Panel

Glare Panel

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

High-Definition and High-Density with Minute Pixels

Enlarged Image

Enlarged Image

Digital

Mammography

3.14 million
pixels

5.24 million
pixels

Pixel Size Comparison

Brightness Stabilization

At startup or upon wakeup, the EIZO patented drift correction
function quickly stabilizes the brightness level. In addition, a sensor
measures the backlight brightness and compensates for brightness
fluctuations caused by the ambient temperature and the passage of time.

Compatibility with Various Graphics Boards

Supports various input signals for compatibility with a wide
r a n g e o f g r a p h i c s b o a r d s s o t h e u s e r c a n s e l e c t t h e b o a r d
a c c o r d i n g t o b u d g e t a n d p e r f o r m a n c e r e q u i r e m e n t s . T h i s
includes single link and dual link signals, as well as packed-
pixel signal for users requiring 12-bit or 10-bit simultaneous
display.

Backlight Saver

With ScreenManager Pro for Medical utility software
installed, the Backlight Saver function allows for the monitor’s
backlight to turn off when the screen saver is activated and
the monitor’s backlight to turn on when the computer comes
out of the screen saver mode. This function helps to reduce
power consumption when the monitor is used for a prolonged
period of time.

High-Resolution for Delicate Grayscale Shadings

When selecting a monitor, it’s important to consider the “information
volume” of the medical images to be displayed.

Provided by Anjo Kosei Hospital

Please note that the reflection of surrounding light may increase when used in a bright environment.
Available with either a glare panel (GS520-CLG, GS520-BLG) or anti-glare panel (GS520-CL, GS520-BL).

This brightness stabilization function is EIZO patented technology (Japan patent numbers 3171808 and
3193315, US patent number 6188380).

0.165mm pixel size is the smallest pixel size among current
EIZO medical monitors providing high-definition and high-
density display without shadings appearing granular.

Mammography

Chest CR

Displaying
only 72%

DUE for Brightness Uniformity

Input Level 80% = 204/255

CCD Camera

Brightness Map

Input Level 80% = 204/255

CCD Camera

Brightness Map

100%

70%

lower

100%

70%

lower

EIZO's confidence in its product quality extends to brightness
stability which is also covered during the usage time specified in the
warranty.

Brightness Stability Within Usage Time Guaranteed

EIZO and its authorized distributors offer a five-year
limited warranty for RadiForce GS520.

Full 5-Year Warranty

Calibration Mode Selection

Selectable with the front panel buttons, the CAL Switch function
allows for various calibration modes of different modalities such as
digital mammography, CR, and CT images. Furthermore, with
ScreenManager Pro for Medical installed, auto mode settings can be
made with the Auto CAL Switch function.

Mammo Mode

Brightness : 500 cd/m

2

Grayscale : DICOM Part 14

Brightness : 450 cd/m

2

Grayscale : DICOM Part 14

Brightness : 350 cd/m

2

Grayscale : γ=2.2

CR Mode

CT/MRI Mode

Without DUE

With DUE

1

Requirement

3

Requirement

5

Requirement

4

Requirement

2

Requirement

To sustain image consistency, brightness uniformity is
required for medical imaging monitors. However, unified
luminance of the backlight is difficult to attain due to the

characteristics of LCD monitors. The Digital Uniformity Equalizer
(DUE) function provides optimum backlight luminance uniformity.

3

4

Distinctive Glare Panel

Featuring high-resolution, high-definition and multi-grayscale, the RadiForce GS520 is

Featuring high-resolution, high-definition and multi-grayscale, the RadiForce GS520 is
designed specifically for digital mammography with a choice of distinctive glare panel to

designed specifically for digital mammography with a choice of distinctive glare panel to
meet diversified environment usage.

meet diversified environment usage.

MONOCHROME
LCD MONITOR

Medical
Images
(Example)

Monitor
Resolution

Glare Panel

Incoming
Light

Reflections

Anti-Glare Panel

Incoming
Light

Reflections

Light from Backlight

Light from Backlight

Change modes

with one touch.

Backlight

Sensor

Backside

of Panel

Hole for

Light

Transmission

Other Manufacturer's Monitor

with Drift Correction

EIZO RadiForce Monitors

with Drift Correction

Seconds

Drift Correction Comparison

Backlights
switch off

automatically

Cuts power consumption
by 11%

18 hours / day

2 hours

16 hours / day

Along with EIZO's new frame rate control technology
(patent pending), 4,096 (12-bit) grayscale tones can be
displayed simultaneously from an abundant palette of

13,771 tones for high-definition digital mammography.

Main Board

Brightness (cd/m

2

)

“Information volume” of a digital mammography image
should exceed 5 million pixels. When a lower resolution
monitor displays this “information volume,” the monitor

stretches the information forcing the mosaic to appear as shadings.
With a 2048 x 2560 resolution or 5.24 million pixels, the stretching
effect is minimized and the mosaic becomes suitable for rendering
subtle masses and calcifications, within the mammography image.

Same Brightness !