Settling time, Analog output output characteristics, Accuracy information – National Instruments NI 7831R User Manual
Page 47
Appendix A
Specifications
© National Instruments Corporation
A-3
Settling Time
Crosstalk................................................. –80 dB, DC to 100 kHz
Analog Output
Output Characteristics
Number of channels ............................... 8 single-ended, voltage output
Resolution .............................................. 16 bits, 1 in 65,536
Update time ............................................ 1.0
µs
Max update rate...................................... 1 MS/s
Type of DAC.......................................... Enhanced R-2R
Data transfers ......................................... Interrupts, programmed I/O
Accuracy Information
Step Size
Accuracy
16 LSB
4 LSB
2 LSB
±20.0 V
7.5
µ
s
10.3
µ
s
40
µ
s
±2.0 V
2.7
µ
s
4.1
µ
s
5.1
µ
s
±0.2 V
1.7
µ
s
2.9
µ
s
3.6
µ
s
Nominal Range (V)
Absolute Accuracy
Absolute
Accuracy at
Full Scale
(mV)
% of Reading
Offset (
µ
V)
Temp Drift
(%/°C)
Positive Full
Scale
Negative Full
Scale
24 Hours
1 Year
10.0
–10.0
0.0335
0.0351
2366
0.0005
5.88
Note: Accuracies are valid for analog output following an internal calibration. Analog output accuracies are listed for
operation temperatures within
±
1 °C of internal calibration temperature and
±
10 °C of external or factory calibration
temperature. Temp Drift applies only if ambient is greater than
±