SCXI-1321 Terminal Block Installation Guide

10

ni.com

Assuming a strain-gauge range with a gauge factor of GF = 2 and a
quarter-bridge configuration, this range corresponds to ±1,498

µε, as given

by the strain formula for a quarter-bridge strain-gauge configuration:

where

Figure 3. Quarter-Bridge Nulling Circuit

ε

4V

r

GF 1 2V

r

+

(

)

-------------------------------

=

V

r

strained voltage static unstrained voltage

V

exc

--------------------------------------------------------------------------------------------------------

=

CH+

CH+

CH–

EX+

EX–

+

+

R

1

R

2

4.5 k

4.5 k

R

SCAL

301 k

R

Trim

10 k

R

Null

39.1 k

(Screw Adjusts Potentiometer)

R

L

R

L

R

L

R

3

(dummy)

R

4

(gauge)

EX+

CH+

CH+

CH–

EX+

EX–

R

Trim

SCXI-1121

SCXI-1321

Transducer

V

EX

Note: R

1

and R

2

are completion resistors.