Figure 3-4. power connector, Table 3-2. power connector terminals, Earth ground connection – National Instruments NI 1450 Series User Manual
Page 38: Earth ground connection -7
![background image](/manuals/190997/38/background.png)
Chapter 3
LEDs, DIP Switches, and Connectors
© National Instruments Corporation
3-7
NI 1450 Series Compact Vision System User Manual
Figure 3-4 shows the power connector on the NI 1450, and Table 3-2
describes each terminal on the connector.
Figure 3-4. Power Connector
Earth Ground Connection
Some system setups may require using the grounding lug on the NI 1450 to
connect the chassis to earth ground. Connecting the grounding lug, shown
in Figure 3-5, to earth ground connects the common of the main power to
earth ground through the NI 1450 chassis.
Note
An earth ground connection does not connect Ciso to earth ground.
Table 3-2. Power Connector Terminals
Terminal
Description
V
Main power (24 VDC ±10%)
C
Common
Viso
Isolated power (5 to 30 VDC)
Ciso
Isolated common
V
iso
(5-30 VDC)
C
C
iso
POWER
V
(24 VDC ±10%)
Viso
(5-30 VDC)
C
V
(24 VDC ±10%)
Ciso
POWER