beautypg.com

Specifications, Monitor (m2922a), Standards compliance – Philips M2925A User Manual

Page 4: Electromagnetic compatibility, Environmental, Dimensions, Power - battery, Power - external power supply, Display, Ultrasound

background image

4

Specifications

Monitor (M2922A)

Federal Law restricts this device to use by or on the
order of a physician.

1

Standards Compliance

The FM-2 Portable Antepartum Fetal Monitor
complies with the requirements of the Medical
Device Directive 93/42/EEC and the following
safety standards:

• EN60601-1/IEC60601-1, EN60601-1-1,

EN60101-2-37

• IEC 1157
• IEC 60601-1
• UL2601-1
• CAN/CSA C22.2#601.1-M90

Class II equipment, double insulated
Continuous operation

Electromagnetic Compatibility

Complies with EN60601-1-2

Environmental

Operating Temp:

-10

o

C to 45

o

C

(50

o

F to 110

o

F)

Storage Temp:

-20

o

C to 60

o

C

(-29

o

F to 175

o

F)

Relative Humidity: 20% to 90% non

condensing

Altitude:

0-3048m (10,000 ft)

Dimensions

Height:

15.24 cm (6.0 inches)

Width:

26.4 cm (10.4 inches)

Depth:

13.7 cm (5.4 inches)

Weight:

2.27 kg (5lb)

Power - Battery

Recharge without

Maximum 8 hours to

monitoring:

full recharge when no
monitoring occurs (typical
4.75 hours).

Low power warning: Approximately 30 minutes

at 25

o

C.

Power - External Power Supply

AC-powered

110-120V~, 50-60Hz,
0.15A
or
220-240V~, 50-60Hz,
0.08A

Display

Backlight:

Cold cathode fluorescent

Active area:

3.78” x 2.83”
(320 x 240 dots)

Resolution:

85 dots per inch

Contrast: 10:1

Ultrasound

I

sata

: <

4mW

BPM range:

Paper scale:

USA paper: 30-240 BPM,
International Paper: 50-210
BPM.

Ultrasound
processing range:

50-240 BPM.

Accuracy:

±

1%

±

1 BPM.

Leakage current:

10

µ

A @ 264 VAC applied

to transducer.

Isolation:

>4 kV RMS, Type BF
applied part.

Toco

Range:

0-100 relative units

Resolution: 1

count

Accuracy:

±

1%

±

1 relative unit

Leakage:

10

µ

A @ 264 VAC applied

to transducer

Isolation:

>4 kV RMS, Type BF
applied part

Transducers

Doppler Ultrasound FHR (M1356A)

Weight:

185 g (6.5 oz)

Cable length:

2.5m (8ft 2in)

Diameter: 75mm

(2.95in)

Latex free

External Toco (M1355A)

Transducer Weight: 180 g (6.3 oz)
Cable length: 2.5m (8ft 2in)
Diameter: 75mm (2.95in)
Latex free

Remote Event Marker (15249A)

Length: 2.8m (9ft 2in)

1.USA only

Parameter

Input

Connector

Resulting

Isolation

US

BF

BF

Toco

BF

BF

Type:

Lead acid, rechargeable.

Operation:

minimum 2-hours fully
charged at 10

o

C, longer

at higher temperature.

Recharge during
monitoring:

Maximum 14 hours full
recharge during monitor-
ing (typical 11 hours).

This manual is related to the following products: