8. Place the new batteries in the power base. Make sure that the terminals are facing inwards. If the battery

terminals are in the middle of the batteries, then make sure that the positive terminal of the rear battery is facing

the negative terminal of the front battery on the left side.

9. Looking at the power base from the front, connect the red wire from the left side to the positive (+) terminal

on the back battery and connect the black wire from the left side to the negative (-) terminal on the front battery.

WARNING! Make sure you tighten the fasteners so that the connections are secure.

10. Connect the red wire from the right side to the positive (+) terminal on the front battery and connect the black

wire from the right side to the negative (-) terminal on the rear battery.

11. Reinstall the shroud.

12. Reinstall the seat.

13. Plug in the connectors to the electronics tray.

When to See Your Authorized Pride Provider for Service
The following symptoms could indicate a serious problem with your power chair. If necessary, contact your

authorized Pride Provider. When calling, have the model number, serial number, nature of the problem, and the

trouble code if available.

„ Motor noise

„ Frayed harnesses

„ Cracked or broken connectors

„ Uneven wear on any of the tires

„ Jerky motion

„ Pulling to one side

„ Bent or broken wheel assemblies

„ Does not power up

„ Powers up, but does not move

„ Loose seat or seating components

Corrective Maintenance
If the battery condition meter does not light up when you turn on the power:

„ Check the harness connections. Make sure they are tight.

„ Check the circuit breaker. Reset it if necessary.

„ Check the battery connections.

If the above conditions prove normal, you can load test the batteries with a battery load tester. These testers are

available at automotive parts stores. Disconnect both batteries before load testing and follow the directions that

come with the load tester. If either one of the batteries fails the load test, replace both of them. If your power chair

still does not power up, contact your authorized Pride Provider.

V I I . C A R E A N D M A I N T E N A N C E

Figure 22. Battery Wiring Diagram

You may need the following to replace your batteries:

„ metric/standard socket set and ratchet

„ adjustable wrench

To replace the batteries:

1. Turn off the power to the controller.

2. Make sure that the power chair is in drive mode. See III.

“Your Power Chair.”

3. Unplug the connectors from the electronics tray.

4. Remove the seat. See V. “Comfort Adjustments.”

5. Remove the shroud.

6. Disconnect the wiring harnesses from the batteries.

7. Remove the batteries from the power base.

This manual is related to the following products: