beautypg.com

Alarm test, Reverse gas flow procedure, Operating instructions – Precision Trading PM5200 User Manual

Page 11

background image

10

Blender

Air-Oxygen

ALARM TEST

1. The Air-Oxygen Blender is installed and the flowmeter is turned

on. Disconnect or turn off the air supply line to the Air-Oxygen

Blender.

2. The Blender should alarm with a loud whistle noise. The whistle

indicates the alarm is operating correctly.

3. Reconnect and activate the air supply line to the Blender, the

alarm should stop whistling.

4. Disconnect or turn off the oxygen supply line to the Blender.

5. The Blender should alarm with a loud whistle noise. The whistle

indicates the alarm is operating correctly.

6. Reconnect and activate the oxygen supply line to the

Blender, the alarm should stop whistling.

7. If alarm fails to function properly, DO NOT USE.

REVERSE gAS FLOW PROCEdURE

(

CE Requirements ONLY)

1. Disconnect the oxygen hose from the gas source. Remove all

outlet connections from the Blender to ensure that there is no

outlet flow.

2. Place the free end of the oxygen supply hose under water. While

gradually increasing the air supply pressure from 30-75 psi (2.07-

5.17 bar) check for leakage past the oxygen inlet check valve.

3. Replace the Duckbill Check Valve in the oxygen inlet if

bubbles indicate leakage. Reference Air-Oxygen Blender

Service manual (P/N 504827.)

4. Repeat steps 1-3 to check for leakage past the air inlet check

valve.

OPERATiNg iNSTRUCTiONS

CAUTiON

Inspect the Air-Oxygen Blender for visual damage before
use, DO NOT USE if damaged.

1. Turn “ON” the Air and Oxygen supply.
2. Adjust the Oxygen Concentration Dial to the prescribed

concentration.

NOTE: The Oxygen Concentration Dial does not rotate

360°.

dO NOT force dial less than 21% or over 100%

oxygen, this will damage the Blender.

This manual is related to the following products: