Component description, Installation – Precision Trading PM5200 User Manual
Page 10
9
Blender
Air-Oxygen
COMPONENT dESCRiPTiON
ITEM
DESCRIPTION
F
Air inlet Fitting
A male DISS or NIST air fitting with one way valve
that is used to connect an air supply hose
.
g
Alarm
An audible alarm that sounds due to an excessive
pressure drop or deletion of either gas supply.
H
Manifold Outlet (Optional)
manifold with 3 primary outlets.
i
Rear Slide Mount with dove tail.
iNSTALLATiON
WARNiNg
Read this User manual before installing or operating the
Air-Oxygen Blender.
Confirm the concentration of air/oxygen with an Oxygen
Analyzer/monitor.
•
•
1. Secure the Air-Oxygen Blender to a wall or pole bracket in an
upright position.
2. It is recommended to install a condensation trap in the air
supply line.
3. Connect the air and oxygen supply lines to the appropriate
inlet fittings on the bottom of the Blender.
4. Attach a flowmeter, or other metering device to one of the
outlet ports.
Primary Outlets Flow capacity:
• High Flow Blender (Pm 5200 model) 15 lpm to 120 lpm
• Low Flow Blender (Pm 5300 model) 3 lpm to 30 lpm
Auxiliary Outlet:
This will bleed off some of the air/oxygen mixture to
maintain concentration accuracy at the Low Flow setting.
• High Flow Blender (Pm 5200 model) 15 lpm or less
• Low Flow Blender (Pm 5300 model) 3 lpm or less
5. Attach a supply line to the outlet port of the flowmeter.