Liebert Process Fluid Chiller PS120A User Manual
Liebert Refrigerators
Process Fluid Chiller Designed for Siemens Magnetom Open 1
Keeping Business in Business
Process Fluid Chiller
Designed for Siemens Magnetom Open
• Dedicated, equipment-matched cooling supplies the proper temperature and water flow year round.
• Factory-tested, piped, wired and charged for hassle-free startup and operation.
• Clean, non-ferrous cooling loop, required for sensitive medical equipment.
• Local service and parts for quick and knowledgeable maintenance and service.
• Optional switch-over to alternate water source provides redundant, continuous operation.
• Optional remote monitoring systems keep you informed of chiller operation.
• UL/CSA listing on standard models.
• ISO 9001 certified company.
Magnetom Open Specifications
Fluid Specified
Water
Heat Dissipation—BTUH (kW)
75,000 (22)
Fluid Temperature Range—°F (°C)
41-48 (5-9)
Fluid Flow Rate—GPM (l/m)
8.5 (32.2)
Fluid Pressure Drop—PSI (kPa)
36.3 (250)
Note: Performance is based on 95°F condenser ambient.
Process Fluid Chiller Specifications
Model Number
PS120A
PS120A
Fluid Circulated
Water
Ethylene Glycol
Net Cooling Capacity—BTUH (kW)
91,600 (26.8)
84,400 (24.7)
Leaving Fluid Temperature—°F (°C)
45 (7.2)
45 (7.2)
Fluid Flow Rate—GPM (l/m)
8.5 (32.2)
10.5 (40)
Magnetom Open Pressure Drop—PSI (kPa)
21.8 (150)
49.3 (340)
Chiller Pressure Drop—PSI (kPa)
3.1 (21.4)
6.8 (46.9)
Pump Total Head—PSI (kPa)
65 (448.1)
64 (441.2)
Head Available for Piping Circuit—PSI (kPa)
40.1 (276)
7.9 (54.8)
Pump HP (kW)
3 (2.2)
3 (2.2)
Based on Siemens specification dated 07/97