Measuring Temperature

Toggle between forehead scan mode and object scan mode by

pressing, holding momentarily, and releasing the MODE button.
Taking a temperature reading from the forehead:

• Press the POWER button and hold it

for one second to turn the device on.

Release it, and the system will power

up, which includes a self test. During

this period the device will momentarily

display all symbols on the LCD (see

Figure 4).

• Press and hold the SCAN button and aim the thermom-

eter at the patient’s forehead, maintaining a distance

of 6 cm ~ 8 cm (2

3

/

8

" ~ 3

1

/

8

") between the Non-Touch

Thermometer and the patient’s forehead. You will hear

a continuous beeping sound, accompanied by a flash-

ing LED. Release the SCAN button; you will hear a short

beep. Do not move the thermometer away from the fore-

head until you hear the beep. Moving the thermometer

away from the forehead before hearing the beep may

result in an inaccurate temperature reading. Please note:

you will not hear a continuous beeping sound unless the

thermometer is the correct distance from the forehead

(see Figure 5). Results will display on the LCD.