Measurement principle – Konica Minolta PULSOX 300 User Manual
Page 30

28
Appendix
■
Spectral Absorption of Hb and HbO
2
Absorption coef
ficient
Wavelength
Red light
Infrared light
Measurement Principle
This instrument measures the oxygen saturation (SpO
2
) in arte-
rial blood and pulse rate continuously in a non-invasive method
by applying optical principles.
by applying optical principles.
With this instrument, SpO
2
is defi ned by the following equation.
where
C (Hb):
Concentration of reduced hemoglobin
C (HbO
2
): Concentration of oxyhemoglobin
The light-absorption characteristics of reduced hemoglobin (Hb)
are very different from those of oxyhemoglobin. This instrument
measures the changes in the absorption of red and infrared
lights passing through the tissue to determine the SpO
2
of the
blood. Thus, this method is free from the effects of skin color,
muscles, bones, and veins.