Appendix
Specification
■
Dual-wavelength pulse type (665nm/880nm) arterial blood oxygen satura-
tion measuring instrument
• Name:
Oxygen Saturation Monitor
• Model name: PULSOX-300
■
Functions:
• Measuring range: SpO
2
:
0 to 100% SpO
2
Pulse rate: 30 to 230 bpm
• Accuracy:
SpO
2
:
±2% SpO
2
(70 to 100% range, 1S.D.)
Pulse rate: ±2 bpm (30 to 100 range)
or ±2% of value (100 to 230 range)
✻
PULSOX-300 has been clinically validated for use on adults.
■
Display
Display type: Liquid crystal display
Oxygen saturation (SpO
2
)
Pulse rate number
Pulse level meter (5 steps)
Battery indication
Error messages indication
■
Warning functions
SpO
2
: Low SpO
2
warning (flashing display)
Battery indication (flashing when the battery is nealy empty)
Error message indication
■
Operating conditions
temperature/humidity range
0 to 40°C (32 to 104°F); 30 to 85% relative humidity with no condensation
atmospheric pressure/altitude range
700 to 1060hPa (altitude: –400 to 3000m)
■
Storage conditions
temperature/humidity range
–10 to 60°C (14 to 140°F); 10 to 95% relative humidity with no condensation
atmospheric pressure/altitude range
700 to 1060hPa (altitude: –400 to 3000m)
■
Power
1 AAA-size battery: 1.5V 100mW (Service life under continual usage:
approx. 16 hours with alkaline battery)
■
Dimensions (W × H × D)
68 Ч 58 Ч 15 mm
■
Weight
56g (including battery/excluding wrist band and probes)
■
Usable life
6 years [Verified by KONICA MINOLTA SENSING, INC. (based on own data)]