Siemens Spectrodensitometer FD-7 User Manual
Page 4
Certificate No : JQA-E-80027
Registration Date : March 12, 1997
Certificate No : LRQ 0960094/A
Registration Date : March 3, 1995
For correct use and for your safety, be sure to read the instruction
manual before using the instrument.
s
Always connect the instrument to the specified power supply
voltage. Improper connection may cause a fire or electric shock.
SAFETY PRECAUTIONS
9242-4897-10 BDEGPK 7
©2010 KONICA MINOLTA, INC.
• Displays shown are for illustration purpose only.
• The specifications and drawings given here are su
bject to change without prior notice.
• KONICA MINOLTA and the Konica Minolta logo and the symbol mark, and "Giving Shape to Ideas" and SpectraMagic™ are
registered trademarks or trademarks of KONICA MINOLTA, INC.
• The basICColor logo is a registered trademark of basICColor GmbH.
• Other company names and product names used herein are trademarks or registered trademarks of their respective companies.
Function
*
New
FD-7
FD-5
Density measurement functions
Density, density difference
Dot area
Dot gain
Trapping
PS plate dot area
PS plate dot gain
Spot color density
Gray balance
Midtone spread
ISO 12647 check
Target match
Colorimetric measurement functions
L a b
L C h
Hunter Lab
Yxy
XYZ
E*
ab
(CIE1976)
E*
94
(CIE1994)
E
00
(CIE2000)
E (Hunter)
CMC (l:c)
Illuminance measurement functions
Illuminance
Correlated color temperature
Paper index
WI/Tint (ASTME313-96)
ISO Brightness (ISO2470-1)
D
65
Brightness (ISO2470-2)
Fluorescence index
Spectral reflectance
Spectral data output
Memory
Target density
30
30
Target color
30
30
Color sets of 15 colors each
3
50
50
Other functions
Manual scan
2
Automatic function
(density, dot area, color)
PASS/FAIL judgment
Software
basICColor catch all
FD-S1w
SpectraMagic NX
ClrChrt
(Included with ColorScoutA3+)
Main specifications
Illumination/viewing
system
45°a: 0°(annular illumination)
1
Conforms to CIE No. 15, ISO 7724/1, DIN5033 Teil 7, ASTM E 1164, and JIS Z
8722 Condition a for reflectance measurements.
Spectral separation
device
Concave grating
Wavelength range
Spectral reflectance: 380 to 730 nm; Spectral irradiance (FD-7 only): 360 to 730 nm
Wavelength pitch
10 nm
Half bandwidth
Approx. 10 nm
Measurement area
Ø3.5 mm
Light source
LED
Measurement range
Density: 0.0D to 2.5D; Reflectance: 0 to 150%
Short-term repeatability
Density:
ı0.01D
Without polarization filter 0.0D ~ 2.5D, Yellow 0.0D ~ 2.0D
With polarization filter
0.0D ~ 2.5D, Yellow 0.0D ~ 1.8D
(When measurements taken 30 times at 10-second intervals after
white calibration has been performed)
Colorimetric: Within
ı E
00
0.05 (Without polarization filter)
(When white plate is measured 30 times at 10-second intervals
after white calibration has been performed)
Inter-instrument
agreement
Within E
00
0.3 (Average of 12 BCRA Series II color tiles compared to values
measured with a master body under Konica Minolta standard conditions; without
polarization filter )
Measurement time
Approx. 1.4 s (single-point reflectance measurement without polarization filter)
Measurement
conditions
Corresponding to ISO 13655 Measurement Conditions M0 (CIE Illuminant A),
M1 (CIE Illuminant D50), M2 (illumination with UV-cut filter), and M3 (M2 +
polarization filter); User-defined illuminant
Illuminants
A, C, D50, D65, ID50, ID65, F2, F6, F7, F8, F9, F10, F11, F12, User-defined
illuminant
Observers
2° Standard Observer, 10° Standard Observer
Density
ISO Status T, ISO Status E, ISO Status A, ISO Status I; DIN16536
Display language
English, French, German, Spanish, Japanese, Chinese (Simplified)
Interface
USB 2.0
Output data
2
Displayed values; Spectral reflectance data (FD-7 only) ; Spectral irradiance data
(FD-7 only)
Power
Rechargeable internal lithium-ion battery (Number of measurements per charge:
Approx. 2,000 when new without polarization filter); AC adapter; USB bus power
Dimensions (W × D × H) 70 Ч 165 Ч 83 mm (Body only); 90 Ч 172 Ч 84 mm (With target mask attached)
Weight
Approx. 350 g (Body only); Approx. 430 g (With target mask attached)
Operating temperature/
humidity range
10 to 35°C, 30 to 85% relative humidity with no condensation
Storage temperature/
humidity range
0 to 45°C, 0 to 85% relative humidity with no condensation
Standard accessories
White Calibration Plate, Target Mask, Protection Glass, AC Adapter, USB Cable,
Soft Case, Polarization Filter, Data Management Software FD-S1w, Illuminance
Adapter (FD-7 only), Ruler (FD-7 only)
Optional accessories
Color Management Software basICColor catch all,
XY Automatic Color Measurement Stage Color Scout A3+,
Color Management Software SpectraMagic NX
1 Illumination for wavelengths under 400nm is unidirectional.
2 Available when using PC software.
3 Used for ISO 12647 Check / Target match; Must be set using included FD-S1w software.