Dehumidification mode – Sharp CV-13NH User Manual
Page 21
ENGLISH
DEHUMIDIFICATION MODE
E-19
In this mode, the air conditioner dehumidifi es the room.
1
Press the MODE button to select DEHU-
MIDIFICATION mode.
COOL
DEHUM
FAN
VENT
2
Press the power button to start operation.
• The red OPERATION lamp on the unit will light.
• The temperature cannot be set.
• The fan speed is preset to AUTO and cannot be
changed.
TO TURN OFF
Press the POWER button again.
• The red OPERATION lamp on the unit will turn off.
If draining of the water is not desirable, install the exhaust hose (See Page 14), turn the drainage nozzle to the
CLOSE position, and check that the drainage nozzle is covered with the stopcock. In this operation, the water
tank inside the unit may become full, the unit will stop operating and the TIMER, OPERATION and MEGA COOL
lamps will blink. In this case, drain out the water within the unit (See Page 26).
Ensure that the stopcock is securely inserted into the lower drainpipe nozzle and that the nozzle has not been
damaged.
1
Remove the exhaust hose (See Page 15)
2
Turn the drainage nozzle to the OPEN position.
3
Pull the stopcock out from the drainage nozzle.
• When the stopcock is removed, a small amount of water
may be discharged from the drainage nozzle.
• Always perform this procedure with the unit turned off.
Drain water will spout out if attempted during operation.
4
Attach the drain grommet to a standard commer-
cially-available hose. (5/8" inner diameter, 7/8"
outer diameter)
5
Attach the hose to the drainage nozzle, and se-
cure it with the hose band.
• Insert the hose securely into a container with a minimum
capacity of 3
1/2
gallons. Be sure to monitor the water
level in the container and empty as necessary. Do not
operate in dehumidify mode for more than 8 hours at a
time. Be sure to empty the water container whenever
dehumidify mode is started. Failure to empty the water
container can cause the container to overfl ow and cause
damage to underlying materials.
• Set the hose sloping downwards for easier drainage.
Moreover, do not bend the hose at any point, nor allow
the end to be submerged in water.
Dehumidifi cation with no container
2
1
Drainage nozzle
Stopcock
Drain grommet
Hose
"OPEN" position
Hose band
CAUTION
When operating dehumidifi cation with container, the unit generates heat during dehumidifi -
cation mode and the room temperature will rise. Operate dehumidifi cation with no container
if you don't want the room temperature to rise. This will help to slightly drop the room tem-
perature, but dehumidifi cation performance will become less effective than when operating
dehumidifi cation with container.
Dehumidifi cation with container