beautypg.com

SSI America BINARY SOLVENT DELIVERY MODULE 90-2581 REV B User Manual

Page 41

background image

A-3

CS OK,x.xx,y,z,PSI,w,v,u/


(x.xxx, xx.xx, or xxx.x)
(y, yy, yyy, or yyyy)
(z, zz, zzz, or zzzz)

Reads the current pump setup, where:

x.xxx, xx.xx, or xxx.x = Flow rate in mL/min

y, yy, yyy, or yyyy = Upper pressure limit

z, zz, zzz, or zzzz = Lower pressure limit

PSI = Units (PSI, ATM, MPA, BAR, or KGC)

w = Pump head size (0 = standard, 1 = macro)

v = Run status (0 = stopped, 1 = running)

u = Pressure Board present = 0; otherwise 1

ID

OK,vx.xx SR3O firmware/

Identifies the pump type and EPROM revision x.xx

UPxxxx

OK/

Sets the upper pressure limit in PSI. The maximum value
for xxxx is 5000 for the plastic head or 6000 for the steel
head; the minimum value is the lower limit plus 100. The
value must be expressed as four digits, i.e., for 900 PSI
xxxx = 0900.

LPxxxx

OK/

Sets the lower pressure limit in PSI. The maximum value
for xxxx is the current upper pressure limit setting minus
100; the minimum value is 0. The value must be
expressed as four digits, i.e., for 100 PSI xxxx = 0100.

SF

OK/

Puts the pump in fault mode. Turns on the FAULT LED
and stops the pump immediately.

RF

OK,x,y,z/

Reads the fault status, where:

x = Motor stall fault (0 = no, 1 = yes)

y = Upper pressure limit fault (0 = no, 1 = yes)

z = Lower pressure limit fault (0 = no, 1 = yes)

KD

OK/

Disables the keypad. (Default status at power-up is
enabled.)

KE

OK/

Enables the keypad.

PCxx

OK/

Sets the pressure compensation value, where xx = the
operating pressure (in PSI divided by 100),
i.e., for 0 PSI xx = 00, for 5000 PSI xx = 50.

RC OK,x/

(x or xx)

Reads the pressure compensation value in hundreds of
PSI, i.e., for 0 PSI x = 0, for 5000 PSI xx = 50.

HTx

OK/

Sets the pump head type, where:

x = 1 for a stainless steel 10 mL/min pump head

x = 2 for a plastic 10 mL/min pump head

x = 3 for a stainless steel 40 mL/min pump head

x = 4 for a plastic 40 mL/min pump head

x = 5 for a stainless steel 5 mL/min pump head
x = 6 for a plastic 5 mL/min pump head
The pump is stopped; and, the pressure compensation
and pressure limits are initialized, when the head type is
changed.

RH

OK,x/

Reads the pump head type, where:

x = 1 for a stainless steel 10 mL/min pump head

x = 2 for a plastic 10 mL/min pump head

x = 3 for a stainless steel 40 mL/min pump head

x = 4 for a plastic 40 mL/min pump head

x = 5 for a stainless steel 5mL/min pump head
x = 6 for a plastic 5 mL/min pump head