SSI America BINARY SOLVENT DELIVERY MODULE 90-2581 REV B User Manual
Page 41

A-3
CS OK,x.xx,y,z,PSI,w,v,u/
 
(x.xxx, xx.xx, or xxx.x) 
(y, yy, yyy, or yyyy) 
(z, zz, zzz, or zzzz) 
Reads the current pump setup, where: 
 
x.xxx, xx.xx, or xxx.x = Flow rate in mL/min
y, yy, yyy, or yyyy = Upper pressure limit
z, zz, zzz, or zzzz = Lower pressure limit
PSI = Units (PSI, ATM, MPA, BAR, or KGC)
w = Pump head size (0 = standard, 1 = macro)
v = Run status (0 = stopped, 1 = running)
u = Pressure Board present = 0; otherwise 1
ID
OK,vx.xx SR3O firmware/
Identifies the pump type and EPROM revision x.xx
UPxxxx
OK/
Sets the upper pressure limit in PSI. The maximum value 
for xxxx is 5000 for the plastic head or 6000 for the steel 
head; the minimum value is the lower limit plus 100. The 
value must be expressed as four digits, i.e., for 900 PSI 
xxxx = 0900. 
LPxxxx
OK/
Sets the lower pressure limit in PSI. The maximum value 
for xxxx is the current upper pressure limit setting minus 
100; the minimum value is 0. The value must be 
expressed as four digits, i.e., for 100 PSI xxxx = 0100. 
SF
OK/
Puts the pump in fault mode. Turns on the FAULT LED 
and stops the pump immediately. 
RF
OK,x,y,z/
Reads the fault status, where: 
 
x = Motor stall fault (0 = no, 1 = yes)
y = Upper pressure limit fault (0 = no, 1 = yes)
z = Lower pressure limit fault (0 = no, 1 = yes)
KD
OK/
Disables the keypad. (Default status at power-up is 
enabled.) 
KE
OK/
Enables the keypad.
PCxx
OK/
Sets the pressure compensation value, where xx = the 
operating pressure (in PSI divided by 100), 
i.e., for 0 PSI xx = 00, for 5000 PSI xx = 50. 
RC OK,x/
(x or xx)
Reads the pressure compensation value in hundreds of 
PSI, i.e., for 0 PSI x = 0, for 5000 PSI xx = 50. 
HTx
OK/
Sets the pump head type, where: 
 
x = 1 for a stainless steel 10 mL/min pump head
x = 2 for a plastic 10 mL/min pump head
x = 3 for a stainless steel 40 mL/min pump head
x = 4 for a plastic 40 mL/min pump head
 x = 5 for a stainless steel 5 mL/min pump head 
 x = 6 for a plastic 5 mL/min pump head 
The pump is stopped; and, the pressure compensation 
and pressure limits are initialized, when the head type is 
changed. 
RH
OK,x/
Reads the pump head type, where: 
 
x = 1 for a stainless steel 10 mL/min pump head
x = 2 for a plastic 10 mL/min pump head
x = 3 for a stainless steel 40 mL/min pump head
x = 4 for a plastic 40 mL/min pump head
 x = 5 for a stainless steel 5mL/min pump head 
 x = 6 for a plastic 5 mL/min pump head 
