Handling batteries, 2 d sp – Siemens 2 D SP User Manual
Page 15
15
2 D SP
Handling batteries
So-called zinc-air batteries are
used today almost exclusively
in hearing instruments to
minimize environmental
damage. These batteries
require oxygen for operation,
which they absorb from the
air. The battery is activated as
soon as the protective foil is pulled off and is used up even
if you do not use your hearing instrument.
You should always pull off the protective foil only if you
actually use the battery. Also, do not leave the batteries
in your hearing instruments if you do not use them over
a longer period. The batteries can leak and damage the
instruments. The batteries keep for from a few days up to
several weeks according to the way they are used. You can
recognize when a change of batteries becomes necessary,
because your hearing instruments become noticeably quieter
towards the end of the battery’s life. Then it is time for new
batteries. Two warning tones sound once per minute, indicating
that you must insert a new battery.
Your hearing instrument specialist can switch off the warning
tones if required.
Zinc-air batteries which are almost completely used up show
the following behaviour: After switching on the instruments
they function completely normally, but then rapidly become
quieter. The more the battery is used up, the faster the gain
falls off. You must then replace such batteries by new ones.
…