Temperature recorder – Sanyo MDF-793AT User Manual
Page 22
21
because of cooling capacity.)
TEMPERATURE RECORDER
The figure below shows the description of a temperature recorder. The temperature recorder is
provided with model AT type.
Feeding of chart
1. Open the door and let down the lever of the penholder; the pen point is apart from the chart. (Fig. 1)
2. Pull the cartridge out of the mounted position. (Fig. 2)
3. Set a new chart in place on the rear bottom of the cartridge. Set the hole on the chart in the cog of the
chart driving assembly and feed the chart in the direction of the arrow by driving the cog wheel.
4. Set the chart according to the mark of day and time. (Fig. 3)
5. When mounting, lay down the cartridge first, and push into the mounting position with the groove set a
the projection on the position.
6. Set up the cartridge and settle in position.
Replacement of cell
Replace the dry cell once a year as follows:
1. Put up the penholder first, then pull out cartridge
from the mounted position.
2. Reset the battery chamber cover with wire, on the
bottom left side, and take out the cell.
3. Set a new cell by turning its anode to this side.
4. Shut the cover after exchanging cells. Mount the
cartridge as before and lay down the penholder.
Recording indicator
Pen holder lever
Door
Warning indicator
Cartridge
Fig. 2
Fig. 3
Fig. 1
Time
Temperature
Cell