Front panel summary – SRS Labs SR510 User Manual
Page 9
5
Front Panel Summary
Signal Inputs
Single Ended (A), True Differential (A-B), or Current (I)
Signal Filters
Bandpass:
Q-of-5 Auto-tracking filter (In or Out)
Line Notch:
Q-of-10 Notch Filter at line frequency (In or Out)
2XLine Notch:
Q-of-10 Notch Filter at twice line frequency (In or Out)
Sensitivity
Full scale sensitivity from 100 nV to 500 mV RMS for voltage inputs
or from 100 fA to 500 nA RMS for current inputs.
Dynamic Reserve
Select Dynamic Reserve
Stability
Sensitivity Ranges
LOW
20 dB
5 ppm
1
µ
V to 500 mV
NORM
40 dB
50 ppm
100 nV to 50 mV
HIGH
60 dB
500 ppm
100 nV to 5 mV
Status Indicators
OVLD Signal Overload
UNLK
PLL is not locked to the reference input
ERR
Illegal or Unrecognized command
ACT
RS232 or GPIB interface Activity
REM
Remote mode: front panel has been locked-out
Display Select
X
Signal Amplitude at the selected phase (AcosØ)
OFST
Display the offset which is being added to the signal output
NOISE Compute and display the noise on the signal
Analog Meters
Displays Signal, Offset, or Noise as a fraction of full scale
Output LCD's
Displays Signal, Offset, or Noise in absolute units
Output BNC's
Output follows Analog Meter, ± 10 V for ± full scale
Expand
Multiplies the Analog Meter and Output voltage by a factor X1 or X10.
REL
Set the Offset to null the output: subsequent readings are relative readings.
Offset
Enables or Disables Offset, and allows any offset (up to full scale) to be entered.
Time Constants
Pre-filter has time constants from 1 mS to 100 S (6 dB/Octave)
Post-filter has time constants of 0, 0.1 or 1.0 S (6 dB/Octave)
ENBW
Equivalent Noise Bandwidth. Specifies the bandwidth when making Noise
measurements. (1Hz or 10 Hz ENBW)
Reference Input
1 M
Ω
Input, 0.5 Hz to 100 KHz, 100 mV minimum
Reference Trigger
Trigger on rising edge, zero crossing, or falling edge
f/2f Mode
PLL can lock to either X1 or X2 of the reference input frequency
Phase Controls
Adjust phase in smoothly accelerating 0.025° steps, or by
90° steps. Press both 90° buttons to zero the phase.
Reference LCD
Display reference phase setting or reference frequency
Power Switch
Instrument settings from the last use are recalled on power-up