beautypg.com

Specifications, Options, Performance data – Sanyo MDF-C2156VAN User Manual

Page 6: Mdf-c2156van, Mdf-c8v mdf-u32v, Mdf-u73v mdf-u53v, Sanyo electric co., ltd. biomedical division, Shr119

background image

Specifications

*Appearance and specifications are subject to change without notice.
*Cooling performance is indicated by the temperature reached at the center of the freezer (at ambient temperature of 30

°C with no load).

In order to use the freezer at a stable temperature for a long time, it is recommended that the temperature be set to at least 5

°C higher

than the indicated lowest temperature. In addition, depending on the usage conditions, it may not be possible to reach the indicated
lowest temperature.

HFC

Model
LCO

2

backup system

LN

2

backup system

Backup system mounting kit
Circular recorder

Chart paper
Ink pen

Slip chart recorder

Chart paper
Ink pen

Recorder mounting bracket

Sensor cover
Storage racks
3 drawers (for lower chamber)
SANYO DAQ (Data Acquisition)
system

MDF-C8V

CVK-UB4

MDF-UBK

MTR-G85

RP-G85

PG-R

MTR-85H

RP-85

DF-38FP

MDF-S3085

(for MTR-85H)

MTR-C8


MTR-480

MTR-2000

MDF-U32V

CVK-UB2

CVK-UBN2

MTR-G85

RP-G85

PG-R

MTR-85H

RP-85

DF-38FP

MDF-S3085

(for MTR-85H)

MDF-30SC

MDF-30R

MTR-480

MTR-2000

MDF-U53V

CVK-UB2

CVK-UBN2

MTR-G85

RP-G85

PG-R

MTR-85H

RP-85

DF-38FP

MDF-S3085

(for MTR-85H)

MDF-70SC

MDF-50R

MTR-480

MTR-2000

MDF-U73V

CVK-UB2

CVK-UBN2

MTR-G85

RP-G85

PG-R

MTR-85H

RP-85

DF-38FP

MDF-S3085

(for MTR-85H)

MDF-70SC

MTR-480

MTR-2000

MDF-C2156VAN

Equipped




MTR-155H

RP-155

DF-38FP

MDF-S30150

(for MTR-155H)



MTR-480

MTR-2000

Options

Distributed by:

We design and manufacture products in accordance with ISO9001 & 14001 requirements.
RoHS (European Restriction of Hazardous Substances directives) compliant

• All data shown here are measured based on our company standards. If detailed

data is required, please contact our company.

• If product inspection records are required, please contact our company. (There

will be a charge for the records.) Please note that the records do not include tem-
perature adjustment range or data on temperature distribution inside the unit.

Performance Data

40

20

0

–20

–120

–40
–60
–80

–100

0

7

6

5

2

8

3

4

1

T

e

mperature (˚C)

T

e

mperature (˚C)

Time (hours)

5

25

20

15

10

30

35

20

10

15

5

T

ime

(

hour

s

)

T

im

e

(hours)

Ambient temperature (˚C)

Ambient temperature (˚C)

0

20

10

30

40

20

25

10

15

5

Time (hours)

40
20

0

–20

–120

–40

–140

–60

–160

–80

–100

0

2

20

4

6

8

10

18

16

12 14

Time (hours)

T

ime (hours)

Ambient temperature (˚C)

T

e

mperature (˚C)

5

25

20

15

10

30

35

20

10

15

5

Ambient temp. and maintenance
of –70

°C by injection of LCO

2

(30kg)

5

25

20

15

10

30

35

20

10

15

5

T

ime (hours)

Ambient temperature (˚C)

5

25

20

15

10

30

35

10

15

20

5

T

ime (hours)

Ambient temperature (˚C)

MDF-C2156VAN

LN

2

backup system

operation characteristics

MDF-C8V

MDF-U32V

Central air temp. in cabinet
Ambient temp. 30

°C, No-load

Ambient temp. and maintenance
of –70

°C by injection of LCO

2

(30kg)

Central air temp. in cabinet
Ambient temp. 30

°C, No-load

Ambient temp. and maintenance
of –70

°C by injection of LCO

2

(30kg)

Central air temp. in cabinet
Ambient temp. 30

°C, No-load

Ambient temp. and maintenance
of –135

°C by injection of LN

2

(50kg)

Central air temp. in cabinet
Ambient temp. 30

°C, No-load

Ambient temp. and maintenance
of –70

°C by injection of LCO

2

(30kg)

Central air temp. in cabinet
Ambient temp. 30

°C, No-load

MDF-U73V

MDF-U53V

T

e

mperature (˚C)

Time (hours)

40
20

0

–20

–120

–40

–140

–60

–160

–80

–100

0

12

6

5

4

3

2

10

1

11

9

8

7

40
20

0

–20

–120

–40

–140

–60

–160

–80

–100

0

12

6

5

4

3

2

10

1

11

9

8

7

Time (hours)

T

e

mperature (˚C)

40
20

0

–20

–120

–40

–140

–60

–160

–80

–100

0

12

6

5

4

3

2

10

1

11

9

8

7

CO

2

backup system

operation characteristics

Pull-up characteristics
during power failure

Pull-down characteristics

Further details regarding our products can be accessed at

http://www.sanyo.co.jp/cmg/biomedical

Model

Temperature range

Maximum cooling performance

Exterior dimensions (W x D x H)

Interior dimensions (W x D x H)

Net weight

Effective capacity

Shelves

Access port

Compressor

Refrigerant

Alarm system

Remote alarm contact

Accessories

Model

Temperature range

Maximum cooling performance

Exterior dimensions (W x D x H)

Interior dimensions (W x D x H)

Net weight

Effective capacity

Access port

Compressor

Refrigerant

Alarm system

Remote alarm contact

Accessories

MDF-C8V

-60 to -80

°C (1°C increments)

-80

°C (Ambient temp. 30°C)

550 x 685 x 945 mm (21.6” x 27.0” x 37.2”)

405 x 490 x 425 mm (15.9” x 19.3” x 16.7”)

Approx. 67 kg (148lbs.)

84 Liters (3.0 cu.ft.)

17 mm diameter, 2 locations backside, bottom

Rotary type, Output: 400W

High/Low temp. alarm (

±5°C to ±20°C),

Power failure alarm, Remote alarm contact:

DC30V, 2A

1 set of keys, 1 scraper

MDF-U73V

1010 x 875 x 2010mm

(39.8” x 34.4” x 79.1”)

870 x 600 x 1400mm

(34.2” x 23.6” x 55.1”)

Approx. 350kg (772 lbs.)

728 Liters (25.7 cu.ft.)

MDF-U53V

–50˚C to –86˚C (1˚C increments)

–86˚C (Ambient temp. 30˚C)

770 x 875 x 1990mm

(30.3” x 34.4” x 78.3”)

630 x 600 x 1380mm

(24.8” x 23.6” x 54.3”)

Approx. 299kg (659 lbs.)

519 Liters (18.3 cu.ft.)

Stainless steel, 3 shelves

17mm diameter, 3 locations (back, bottom left/right corners)

Hermetic type, Output:

1100 W (high stage side), 1100 W (low stage side)

High/low temperature, Power failure, Filter check,

Self diagnostics, Door check,

Part replacement nortification

Allowable contact capacity: DC 30V, 2A

1 set of keys, 1 scraper

MDF-U32V

670 x 867 x 1860mm

(26.4” x 34.1” x 73.2”)

490 x 600 x 1140mm

(19.3” x 23.6” x 44.9”)

Approx. 258kg (569 lbs.)

333 Liters (11.8 cu.ft.)

Hermetic type, Output:

400 W (high stage side),

600 W (low stage side)

Temperature, Power failure,

Filter check, Self diagnostics,

Remote alarm contact

MDF-C2156VAN

-125

°C to -150°C (1°C increments)

-150

°C (Ambient Temp. 30°C)

1730 x 765 x 1010 mm (68.1” x 30.1” x 39.8”)

760 x 495 x 615 mm (29.9” x 19.5” x 24.2”)

Approx. 319 kg (703lbs.)

231 Liters (8.2 cu.ft.)

40 mm diameter, 1 location

Hermetic type, Output: 1100W (high stage

side), 1100W (low stage side)

HFC

High/Low temperature, Power failure,

Filter check, Self diagnosis, Door check,

Part replacement notification

Allowable contact capacity, DC 30V, 2A

1 set of keys, 1 scraper, Connecting tube for

LN

2

backup system

SANYO Electric Co., Ltd.
Biomedical Division

©2008 SANYO Printed in Japan 2008.3 MA

SHR119