Technical considerations – SMC Networks Invivo HNC-63-INT User Manual
Page 8
8
Rev 05
The coil should only be used with the accessories specified in the operator’s
manual.
The use of accessories other than those specified in the operator’s manual
may result in decreased ESD immunity of the coil or MR system, causing
damage to the coil and/or system.
The equipment should not be used with other coils or equipment present in
the MR scanner except as specified in the Operator’s Manual.
Tampering with the cable pins and connector may damage the connector
and affect coil or system performance. Please verify that connector and
pins are not damaged before use.
The coil and accessories require special conditions regarding
electromagnetic compatibility. The coil must be installed and used in a
shielded scan room provided with the MR magnet and system. The user
must ensure that the scan room door is closed during system use. Failure
to do so may cause reciprocal interference with portable and mobile RF
communications equipment, affecting the performance of the MR coil and/
or such equipment.
Technical
Considerations