SMC Networks Invivo HNC-63-INT User Manual
Page 17

17
Rev 05
Run an SNR CHECK scan using the parameters listed below.
TIP: Save this SNR CHECK scan for future QA scans.
Position slices for the SNR CHECK as shown on page 15.
Geometry
Coil Selection
3IB-MRIDevic
-channel Combination
2345
-connection
d
Homogeneity correction none
FOV (mm)
400.00
RFOV (%)
100.00
Foldover suppression
no
Matrix scan
256
-reconstruction
256
Scan percentage (%)
100.00
Stacks
1
-type
parallel
-slices
3
-slice thickness (mm)
5.00
-slice gap
user defined
- - gap (mm)
18.00
-slice orientation
sagittal
-foldover direction
AP
Slice scan order
default
PlanAlign
no
REST slabs
0
Patient position
head first
-orientation
supine
Contrast
Scan mode
MS
-technique
FFE
Contrast enhancement
no
Fast Imaging mode
none
Echoes
1
-partial echo
yes
TE
user defined
-(ms)
15.00
Flip angle (deg)
30.00
TR
user defined
-(ms)
200.00
Half Scan
no
Water fat shift
user defined
-(pixels)
2.000
Shim
no
SPIR
no
MTC
no
SAR mode
default
Gradient mode
regular
Motion
Cardiac synchronisation
no
Respiratory compensation
no
Flow compensation
yes
NSA
2
SMART
no
Dyn/ang
Angio
no
Quantitative flow
no
Manual start
no
Dynamic study
no
Postproc
Preparation phases
full
MIP/MPR
no
Images:
O:M no no no
Autoview image
M
Reference tissue
Grey matter
Preset window contrast
soft
Reconstruction mode
immediate
***Save raw data***
yes
Push to workstation
no
Hardcopy protocol
no
Ringing filter
no
Silent mode
no
Offc/ang
Stacks
1
Stack Offc.AP (P=+mm)
-20.00
RL (L=+mm)
0.00
FH (H=+mm)
0.00
Ang. AP (deg)
0.00
RL (deg)
-0.00
FH (deg)
-0.00
***IMPORTANT: Be certain to SAVE RAW DATA under Post Processing.***