TYPICAL PRODUCT PARAMETERS

HomeFill

®

Compressor

8

Part No 1100873

Input Pressure Required:

14 - 21 psi

Input Flow Required:

2 L/min

Input O

2

% Required:

>90% O

2

Width:

20 ¼ inches

Height:

15 inches

Depth:

16 inches

Weight:

33 lbs

Shipping Weight:

39 lbs

No AP/APG:

Not suitable for use in the presence of a
flammable anaesthetic mixture.

Electrical:

No extension cords.

Placement:

No closer than 7.5 cm (3 in) from any
wall, furniture, draperies, or furniture to
assure sufficient air flow. Avoid deep pile
carpets and heaters, radiators or hot air
registers. Floor location only.

Tubing:

2 m (7 ft) cannula with a maximum 15 m
(50 ft) of Crush-Proof Tubing (DO NOT
pinch).

Time of Operation:

Up to 24 hours per day.

Recommended Storage and
Shipping Temperature:

-29°C to 65°C (-20°F to 150°F) at
15-95% relative humidity.

Environment:

Smoke and soot-free. No confined
spaces (Example: No closets).

Minimum Operating Time:

30 Minutes