TYPICAL PRODUCT PARAMETERS
HomeFill
®
Compressor
8
Part No 1100873
Input Pressure Required:
14 - 21 psi
Input Flow Required:
2 L/min
Input O
2
% Required:
>90% O
2
Width:
20 ¼ inches
Height:
15 inches
Depth:
16 inches
Weight:
33 lbs
Shipping Weight:
39 lbs
No AP/APG:
Not suitable for use in the presence of a
flammable anaesthetic mixture.
Electrical:
No extension cords.
Placement:
No closer than 7.5 cm (3 in) from any
wall, furniture, draperies, or furniture to
assure sufficient air flow. Avoid deep pile
carpets and heaters, radiators or hot air
registers. Floor location only.
Tubing:
2 m (7 ft) cannula with a maximum 15 m
(50 ft) of Crush-Proof Tubing (DO NOT
pinch).
Time of Operation:
Up to 24 hours per day.
Recommended Storage and
Shipping Temperature:
-29°C to 65°C (-20°F to 150°F) at
15-95% relative humidity.
Environment:
Smoke and soot-free. No confined
spaces (Example: No closets).
Minimum Operating Time:
30 Minutes