Indicator light color compressor status effect – Invacare Compressor User Manual
Page 30

SECTION 6—INDICATOR LIGHTS
HomeFill
®
Compressor
30
Part No 1100873
SECTION 6—INDICATOR LIGHTS
INDICATOR LIGHT
COLOR
COMPRESSOR
STATUS
EFFECT
None
Compressor is Off.
O
2
flowing to user if
concentrator is on.
O
2
BELOW NORMAL -
YELLOW
Compressor is On and
warming up
(0-3 minutes after
power is On).
Cylinder filling not
started.
O
2
flowing to user if
concentrator is On.
O
2
from concentrator is
< 90% (any time after
three minute warmup).
Cylinder filling stops, will
restart when O
2
> 90%.
O
2
flowing to user if
concentrator is On.
FILLING - GREEN
Compressor is On and
cylinder is filling. O
2
from concentrator is
>90% (any time after
three minute warmup).
Cylinder is filling.
O
2
flowing to user if
concentrator is On.
FULL - GREEN
Compressor is On. Cyl-
inder is finished
filling (full) -remove
cylinder (FULL light is
On in >10 mins.)
or
Cylinder is not attached
properly - check con-
nection (FULL light is
On in <10 mins.).
Cylinder filling stops.
Cannot fill another
cylinder until
compressor is turned Off
and On again.
O
2
flowing to user if
concentrator is On.
ATTENTION - RED
(Audible Alarm)
Compressor is On.
Cylinder NOT filling
Causes:
Full cylinder attached
when unit turned On -
replace with empty
cylinder
or
Internal failure -
contact dealer.
Compressor does not
begin to fill or stops
filling.
Unit must be turned Off
and On before
filling can occur.
O
2
flowing to user if
concentrator is On.