BLOCK 0

HEX

DEC

PARAMETER

DESCRIPTION

READ/WRITE

source value is based on the selection in
ASRC upper byte where

14

20

SPARE

SPARE

READ/WRITE

15

21

SPARE

SPARE

READ/WRITE

16

22

SPARE

SPARE

READ/WRITE

17

23

TEMPFIL

Temperature Input Filter in seconds
Range = 0 to 3276. The higher the number
the faster the reading update.
DEFAULT = 1000

READ/WRITE

BLOCK 1

HEX

DEC

PARAMETER

DESCRIPTION

READ/WRITE

18

24

MVFIL

Millivolt Input Filter in seconds
Range = 0 to 3276. The higher the number
the faster the reading update.
DEFAULT = 1000

READ/WRITE

19

25

AZERO

LINEAR OFFSET, Y INTERCEPT LINEAR
SCALING FOR INPUT A

READ/WRITE

1A

26

ANUM

LINEAR SPAN VALUE FOR INPUT A

READ/WRITE

1B

27

BZERO

LINEAR OFFSET, Y INTERCEPT LINEAR
SCALING FOR INPUT B

READ/WRITE

1C

14

BNUM

LINEAR SPAN VALUE FOR INPUT B

READ/WRITE

1D

15

PROC

This value is the calculated process value
shown as an integer. The decimal point and
exponent values are required to determine the
actual scaled value.
Range = -999 to 9999.
For example: If the process = oxygen, display
decimal point = 2, and exponent = 6, and
PROC = 1234, then the actual value and
displayed as 12.34 ppm.

READ ONLY

1E

16

COLDJCT

COLD JUNCTION
Where 1 COUNT = 1°F (°C), RANGE = -99 TO
255°F (°C). Note this parameter is an
unsigned integer.

READ ONLY

1F

17

TEMP

MEASURED TEMPERATURE
Where temperature is presented in degrees C
or F, based on the C/F setting. Note this
parameter is an unsigned integer of
temperature -2721 = 62815
Range = max / min range of selected
thermocouple.

READ ONLY

20

18

MV

MEASURED MILLIVOLT
Where this value is scaled in 0.1 mV
increments, i.e. 10001 = 1000.1.
Range = 0 to 2000 mV.

READ ONLY

21

19

DACV1

ANALOG OUTPUT 1
0 to 4095 is 4 to 20 mA In dual mode 4mA = -
100, 12mA = 0, 20mA = +100

READ/WRITE