0 troubleshooting, 0 specifications, control unit – Graham Field ALTADYNE 753004 User Manual

Page 9

background image

9

Lumex

®

AltaDyne AutoFloat

Troubleshooting/Specifications

14.0 Troubleshooting

SYMPTOM:

CORRECTIVE ACTION

Mattress does not

inflate or is too soft.

Make certain the air hoses are not kinked, cracked, or split. Verify the hose couplings

are fully engaged to controller. Verify that all 20 air cells are connected to the air manifold

within mattress.

Power loss indicator

flashing and audible

alarm sounding

Power has been disconnected from unit while in an operational mode.

Restore power, put unit in STANDBY mode, then remove power. Unit must always be

returned to STANDBY mode before disconnecting power.

Indicator adjacent to

ALARM SILENCE

button is flashing and

audible alarm is

sounding.

If in AUTO mode, check the connection of the Active Sensor mattress sensor cord to the

jack on the side of the unit. Reconnect if necessary.

Verify hose couplings are fully engaged to controller. Verify that all 20 air cells are

connected to the air manifold within mattress.

15.0 Specifications, Control Unit

Enclosure Dimensions

10" x 12" x 4

1

/

2

", (25 cm x 30 cm x 11 cm)

Weight

9.5 pounds (4.3 kg)

Power Cord

Detachable 14' , #18 AWG minimum, with ground wire

Overcurrent Protection

Primary 2.5A, 250V, T

Input

120 -230 VAC, 50/60Hz, 120V – 0.44A, 230V – 0.33A

Operating Ambient
Temperature Range

60 to 85°F (16 to 29°C)

Classification

Class I grounded equipment not suitable for use in the presence of a flammable

anesthetic mixture with air or with oxygen or nitrous oxide.

Type BF equipment

UL2601-1, CSA C22.2 NO. 601.1

EN 60601-1

IPXO, enclosed equipment without protection

against ingress of water.

Continuous operation

Electromagnetic
Compatibility

Meets EN60601-1-2:2001

(CISPR 11 Classified as Class B, Group 1 ISM equipment)