beautypg.com

Gilson 508 User Manual

Page 13

background image

Gilson 508 Interface Module User’s Guide

2-3

Installation

2

Elec

tric

al C

o

nnec

tions

Refer to the diagram below to connect the 508 to another Gilson instrument
using a GSIOC cable.

With additional GSIOC cables, you can connect as many as 32 devices. To connect
additional GSIOC cables, connect the male connector to the next cable’s female
connector (use the female connector that’s not joined to the male connector).
Then connect the remaining female connector to the GSIOC port of the device.

Cables should be arranged in a linear fashion. Any “Y-branching” from the main
GSIOC cable may increase noise.

Power

After you’ve made all the necessary connections, attach the power connector to
the POWER input of the interface module and the transformer to a wall outlet.
Since the interface module is designed to be left on at all times, it does not have
a power switch. If you prefer to have a power switch, a switched power outlet
strip is recommended.

The Power light indicates that power is being supplied to the unit.

Euro plug adapter

Locate the Euro Plug Adapter (part number 594505121) supplied with the 508
Interface Module. If necessary, attach to the power cord to connect the interface
module to an AC power source.

9-pin female
connector; connect
to the GSIOC port
on the interface
module

9-pin male connector;
connect to another
GSIOC cable for linking
additional GSIOC devices

9-pin female
connector;
connect to
GSIOC port of
GSIOC device