RS 232 Interface

The RS 232 port is used to connect the SECUTEST

®

PSI module (acces-

sory), which can be installed to the lid of the SECUTEST

®

SII, to a PC or a

barcode scanner.

17.1 Transmission of Measurement Results to the SECUTEST

®

PSI

Test results – except for results from individual measurements (selector
switch in MENU position) and the function test – can be transmitted from
the SECUTEST

®

SII to the SECUTEST

®

PSI module, where they can be

stored and printed out in the form of measuring, test and statistical
reports at any time.

17.2 PC Connection

Connection to an IBM compatible PC is also possible. The PC is
connected to the interface at the test instrument, or to the interface port
at a previously installed SECUTEST

®

PSI module.

17.2.1 Software Evaluation of Measurement Results

Convenient software programs such as PC.doc-WORD, ACCESS or PS3
allow for easy generation of measuring and test reports, as well as
archiving of measured data.

17.2.2 Instrument Control via Interface Commands

All key functions included with the SECUTEST

®

SII can be simulated with

the help of interface protocols, and the following parameters can be
queried:

Type of measurement and measuring range

Test setup

Measurement sequence progress

Detailed measurement results

17.3 Interface Configuration and Protocol

The interface included with the SECUTEST

®

SII is in compliance with the

RS 232 standard.

Technical Data:

Baud Rate

9600 baud, permanently set

Character Length

8 bits

Parity

none

Stop Bits

1

Data Protocol

per DIN 19244, X_ON / X_OFF protocol

Connector Pin Assignments for 9-Pin Subminiature Socket Connector:

1: External in + (for internal use only)
2: TXD (transmitter output)
3: RXD (receiver input)
4: External in +
5: GROUND
6: +5 V (500 mA output, for barcode scanner only)
7: Ext. in –
8: Control output
9: +9 V (1.5 A output, for PSI module only)

9

8

7

6

5

4

3

2

1