beautypg.com

3 table: leakage current types, Table: leakage current types, List of possible options for secutest – Gossen MetraWatt 702 User Manual

Page 7

background image

GOSSEN METRAWATT GMBH

7

1.3

Table: Leakage Current Types

Key

NC = Normal conditions
PAT = Patient application part
PE = Potential earthing

,

mains protective conductor

DPE = Protective conductor of the device under test

1.4

List of Possible Options for SECUTEST

®

SII Device Series

Example for complete type designation (article number) of a
SECUTEST

®

SII:

M7030 E01 KD01 KE01 (only the designation of the basic device, M7030,
and features other than 00 are specified)

1)

Each measured value is documented in this case, as opposed to the results of a
test sequence for which the poorest value for each given test is displayed.

2)

Adapter set for international use (feature B00 - earthing contact socket - included)

3)

With configuration for hospital beds of Safety Class 2 (pre-adjusted)

4)

For multimeter functions, for measurements with accessories and for devices
under test without plug, including 3 plug-on quick clips

DIN VDE
0701-1

DIN VDE
0702-1

DIN VDE
0751-1 (2001)

The following
is measured:

Equivalent
leakage
current

Equivalent
leakage
current

PROBE
(connected to
protective
conductor) to
L & N

Equivalent
device leakage
current

PROBE (open
protective
conductor) to
L & N

Contact
current / mea-
surement for
absence of
voltage

Contact
current / mea-
surement for
absence of
voltage

Probe to PE

Protective con-
ductor to PE

Device leak-
age current
during opera-
tion

Protective con-
ductor inter-
rupted, probe
to PE

Protective
conductor
current with
differential
current
measurement

Protective
conductor
current with
differential
current
measurement

Device leak-
age current
during opera-
tion,
differential cur-
rent method

See chapter
9.3.2

Features

00

01

02

03

04

05

06

07

08

09

10

11

99

Design

A

GM

Mains Connection for
Respective Country

B

D

F

CH

2)

User Interface
Language

C

D

GB

Configuration
(settings in the
setup menu)

D

GM

3)

Printer Module
SECUTEST

®

PSI

E

w/o

with

Additional
Test Sockets
(cannot be retrofitted)

F

w/o

with

4)

Database
DBmed Option
(Z853H)

KB

w/o

with

Remote Control
SK5 Special Cable
(Z745K)

KD

w/o

with

Direct Printing After
Each Measurement for
Automatic Test
Sequences

1)

Read-Out via RS 232,
SECU-dd Option
(Z853L)

KE

w/o

with

DKD calibration
certificate

L

w/o

with

This manual is related to the following products: