11.8 Testing Multiple Outlets for VDE 0702 (optional EL1 adapter)
➭ up to version 5.10:
Do not select “WITH EL1” in the initial menu. The following must
appear in the line: Ext. Cable – WITH EL1. “Manual sequence” must be
selected.
➭ from version 5.11 onwards:
Select “X
WITH EL1” in the initial menu. The following must
appear in the line: Ext. Cable X WITH EL1. “Manual sequence” must be
selected.
➭ A visual inspection must always be performed. It may thus be
necessary to unreel the cable from its drum or reel.
➭ Protective conductor resistance measurement: Contact the first outlet
with the EL1 adapter. Each time you are ready to contact the next
outlet, press the
key to repeat the test.
➭ After measurement has been completed for all outlets, reinsert the test
probe in order to continue with the performance of subsequent tests
as described in chapter 11.6.