5 basic operating equations, 1 charge, 2 current – Fluke TRACKERTM 90100 User Manual
Page 34: 3 absolute dose, 4 absolute dose rate, 5 percent-of-center

Model 90100
Operators Manual
3-6
Pressure Temperature
mm Hg = inch Hg ∗ 25.4
K = (5/9) ∗ (°F -32) - 273.15
mm Hg = 760 ∗ (54,041 - feet ASL) / (54,041 + feet ASL)
K = °C + 273.15
mm Hg = 760 ∗ (16,472 - meters ASL) / (16,472 + meters ASL)
mm Hg = mB ∗ 0.75006
mm Hg = hPa ∗ 0.75006
3.5 Basic Operating Equations
This section lists the basic operating equations used within the 35360A TRACKER Display to produce its
dose and dose rate measurement values for display on the VFD. These equations are expressed in
terms of a set of basic operating quantities which are described in Table 3-3.
3.5.1 Charge
Although charge is not a value which can be displayed directly by the 35360A TRACKER Display, it is an
intermediate result upon which the dose value directly depends.
[nC/count]
QG
)
[counts]
ADZ
-
[counts]
(ADC
=
[nC]
Q
×
3.5.2 Current
Although current is not a value which can be displayed directly by the 35360A TRACKER Display, it is an
intermediate result upon which the dose rate value directly depends.
count]
[(nC/Min)/
IG
[counts])
ADZ
-
[counts]
(ADC
=
[nC/Min]
I
×
3.5.3 Absolute Dose
FPGF
ADCF
Units]
Cal
Chamber
Ion
Units/
[Display
DUCF
Units/nC]
Cal
Chamber
[Ion
ICCF
[nC]
Q
=
Units]
[Display
Display
Dose
Absolute
Ч
Ч
Ч
Ч
3.5.4 Absolute Dose Rate
FPGF
ADCF
Units]
Cal
Chamber
Ion
Units/
[Display
DUCF
Units/nC]
Cal
Chamber
Ion
[
ICCF
[nC/Min]
I
=
Units/Min]
[Display
Display
Rate
Absolute
Ч
Ч
Ч
Ч
3.5.5 Percent-of-Center
Channel
Center
for
Display
Rate
Absolute
Channel
Perhiperal
for
Display
Rate
Absolute
[%]
100
=
[%]
Channel
Perhiperal
for
Display
Percentage
Ч