Specifications – Fluke PN 4060712 User Manual
Page 10

62 MAX/62 MAX +
Users Manual
6
Specifications
62 MAX
62 MAX +
Temperature Range
-30
°C to 500 °C
(-22
°F to 932 °F)
-30
°C to 650 °C
(-22
°F to 1202 °F)
Accuracy
≥0 °C: ±1.5 °C or ±1.5 % of
reading, whichever is
greater
(
≥32 °F: ±3 °F or ±1.5 % of
reading, whichever is
greater)
≥ -10 °C to <0 °C: ±2 °C
(
≥14 °F to <32 °F: ±4 °F)
< -10
°C: ±3 °C
(<14
°F: ±6 °F)
≥0 °C: ±1 °C or ±1 % of
reading, whichever is
greater (
≥32 °F: ±2 °F or
±1 % of reading, whichever
is greater)
≥ -10 °C to <0 °C: ±2 °C
(
≥14 °F to <32 °F: ±4 °F)
< -10
°C: ±3 °C
(<14
°F: ±6 °F)
Response Time (95 %)
<500 ms (95 % of reading)
<300 ms (95 % of reading)
Spectral Response
8 to 14 microns
Emissivity
0.10 to 1.00