Specifications – Fluke 80T-150UA User Manual
Page 2
Specifications
The 80T-150UA will achieve rated accuracy when it is used with any
0.25 % DMM that has an input impedance of
≥
1 M
Ω
.
Electrical
Measurement Range: (See Probe Limitations)
-50 to +150
°
C
-58 to +302
°
F
Accuracy: (Applies for one year after purchase or calibration)
AMBIENT
°
C ACCURACY
+15 to +35
°
C
±
1
°
C from 0 to +100
°
C, decreasing
linearly to
±
3
°
C at -50 and +150
°
C
0 to +15
°
C and
±
2
°
C from 0 to +100
°
C,
+35 to +50
°
C
decreasing linearly to
±
4
°
C at -50
and
+150
°
C
AMBIENT
°
F ACCURACY
+59 to +95
°
F
±
1.8
°
F from +32 to +212
°
F,
decreasing linearly to
±
5.4
°
F at -58
and +302
°
F
+32 to +59
°
F and
±
3.6
°
F from +32 to +212
°
F,
+95 to +122
°
F
decreasing linearly to
±
7.2
°
F at -58
and +302
°
F
Sensitivity (80T-150UA output): 1 mV dc /
°
C or
°
F
Voltage Standoff: 60 V dc or 30 V ac rms (42.4 V peak)
Settling Time: 9.5 seconds to settle within 2
°
for a 50
°
change
RF Field: The 80T-150UA product is unspecified in environments with
EMC fields greater than 100 mV/meter. ESD performance is specified as
Criteria C per EN61326.
Environmental
Ambient Operating Range for Temperature Module: 0 to +50
°
C
(+32 to +122
°
F)
Maximum Temperature Probe Body and Cable: +70
°
C
(+160
°
F) See Probe Limitations
Minimum Temperature Probe Body and Cable:
Minimum Operating Temperature for Cable: -12
°
C (+10
°
F)
Minimum Operating Temperature for Probe Body: -40
°
C (-40
°
F)
See Probe Limitations
Storage Temperature for Unit: -40 to +70
°
C (-40 to +160
°
F)
Humidity: 0 % to 90 % (0
°
C to +35
°
C)
0 % to 70 % (+35
°
C to +50
°
C) (noncondensing)
Altitude: Operating:
≤
3048 m (
≤
10,000 ft)
Storage:
≤
15240 m (
≤
50,000 ft)
Application Force: 15 pounds maximum (probe tip to measured surface)