Ventricular Fibrillation ......................................... Coarse and fine
Monomorphic Ventricular Tachycardia ............... 120 to 300 (by 5) BPM
Polymorphic Ventricular Tachycardia ................. 5 types
Asystole .............................................................. Flat line
ECG Waves
ECG General
Lead configuration .............................................. 12-lead simulation. RA, LL, LA, RL, V1-6 with independent outputs
Lead to lead impedance ..................................... 1000 Ω (nominal)
Rate accuracy.....................................................
±1 % of nominal
ECG Amplitudes
Reference Lead...................................................Selectable, Lead II (default) or Lead I
Settings................................................................0.05 to 0.45 (by 0.05) mV
0.5 to 5.0 (by 0.5) mV
Accuracy (All Performance waves and Normal Sinus R waves)
Lead II..................................................................±2 %
All other leads......................................................±5 %
Defib paddles.......................................................±5 %
Amplitude of ECG signals relative to amplitude setting (in percent)
Lead II reference
Performance waves and R wave detection
Lead
# I II III V1 V2 V3 V4 V5 V6
Ref. Amp.
70 %
100 %
30 %
100 %
100 %
100 %
100 %
100 %
100 %