Change #1 – Fluke IMPULSE 7000DP User Manual
Page 2

Impulse 6000D/7000DP Users
Manual Supplement
4/08 1
Change #1,
46470, 46697, 47714, 48007, 49680
Replace pages 31 through 35, with the following:
Defibrillator Analyzer Specifications
Energy Output Measurement
Compatible Defibrillator Waveshapes
................ Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic
Note
AC Pulsed Biphasic waveform has not been approved in the United States.
Autoranged Measurement
.................................... 0.1 to 600 J
Accuracy
0.1 to 360 J .....................................................
±(1 % of reading + 0.1 J)
360 to 600 J ....................................................
±(1 % of reading + 0.1 J), typical
Note
For Pulsed Bi-Phasic defibrillator, specified accuracy is
±(1.5 % of reading + 0.3 J) on both ranges.
Load resistance
Resistance .......................................................... 50
Ω
Accuracy .........................................................
±1 %, non-inductive (<2 μH)
Pulse trigger level
................................................. 20 V
Pulse width
Range ................................................................. 1.0 to 50.0 ms