Installation
4.1
Lead Wire Identification
The 5623A PRT is equipped with a four-wire cable (see Figure 1). Four lead
wires are used to cancel lead wire resistance. For best results, the readout de-
vice should be equipped to handle four-terminal resistors.
The lead wires are two different colors. Lead wire pairs attached to each end of
the sensor are identified by red and white insulation.
5623A
9
4 Installation
White
Red
White
Red
Figure 1 PRT Schematic