Electrical (cont.)

Measurement
Accuracy, T1, T2,
or T1-T2
(Model 54)

J-, K-, T-, E-, and N-type:

±

[0.05

%

of

reading

+

0.3

o

C (0.5

o

F)] [below -100

o

C

(-148

o

F): add 0.15 % of reading for J-,

K-, E-, and N-type; and 0.45 % of reading
for T-type] R- and S-type:

±

[0.05

%

of

reading

+

0.4

o

C (0.7

o

F)]

Temperature
Coefficient

0.01 % reading

+

0.03

o

C per

o

C

(0.05

o

F per

o

F) outside the specified

+

18

o

C to 28

o

C (

+

64

o

F to

+

82

o

F) range

[below -100

o

C -148

o

F): add 0.04 % of

reading for J-, K-, E-, and N-type; and
0.08 % of reading for T-type]

Electromagnetic
Compatibility

Susceptibility:

±

2

o

C (

±

3.6

o

F) for 80 MHz

to 200 MHz in 1.5 V/m field, for 200 MHz
to 1000 MHz in 3 V/m field.
Emmisions: Commercial Limits per
EN50081-1

Maximum
Differential
Common Mode
Voltage

1 V (Maximum voltage difference between
T1 and T2)

Temperature
Scale

ITS-90

Applicable
Standards

NIST-175

Accuracy is specified for ambient temperatures between 18

o

C

(64

o

F) and 28

o

C (82

o

F) for a period of 1 year. The above

specifications do not include thermocouple error.

Replacement Parts and Accessories

Accessory Part

Number

Holster and Flex Stand

 Assembly

1272438

AA NEDA 15A IEC LR6 batteries

376756

80PK-1 K-Type Bead Thermocouple

773135

CD-ROM 1276106

Service Manual

1276123

FlukeView Forms

FVF-SC2