Controller operation

In normal operation the controller displays cabinet temperatures
in degrees C (default) or user-selected degrees F. Degrees C
temperatures are displayed to 1 decimal point.

The controller is pre-programmed with a +3.5 C (+38 F) set point
which provides a compressor cut-in at +3.5 C (+38 F) and cut-out at
+5.0 C (+41 F).

All set points have a 1.5 C differential. The 1.5 C differential means
that with a +3.5 C set point, for example, the compressor will turn off
at +3.5 C (+38 F) and turn on when the cabinet temperature reaches
+5 C (+41 F).

This temperature may not refl ect the temperature of the blood
products in the refrigerator. The temperature of the blood products will be refl ected on the recorder chart and the
recorder's digital temperature display. At times there may be a difference between the two displayed temperatures.
Adjust the controller SET value to achieve your desired blood product temperature as displayed on your chart
recorder.

5

2.5

°C

refrigeration
system
energized

Fig. 5

To display temperature cut-out

STEP INPUT

DISPLAY

1

Press and release SET

Current cut-out temperature will display for approximately 5 seconds.

Display will return to current refrigerator temperature.

To change temperature cut-out

STEP INPUT

DISPLAY

1

Press and hold SET for

Current cut-out temperature displayed and °C will fl ash

3

seconds

2

Press UP or DOWN arrows to

New cut-out temperature displayed

desired cut-out temperature

3

Press and release SET

New cut-out temperature blinks three times, then current refrigerator

temperature will display

Refer to the chart below for the set point for your application's required temperature range.

SET POINT DEGREES C

CUT-IN DEGREES C

CUT-OUT DEGREES C

NORMAL DISPLAY RANGE DEGREES C

+3.5 C

+5.0 C

+3.5 C

1.5 to 5.0 C