The key to eliminating an error is knowing its source and type along with its polarity and

magnitude. Generally, the source is simple to detect and is represented by the specific

parameter. The type is usually a function of the instrument’s design and manufacturing

process. Within a given instrument, an error can be either random or systematic, as well

as random or systematic within the instrument’s population. The user is free to consult

Fluke for recommended methods of minimizing error source contributions.

In summation, total error can and should be managed by the control of the three general
error sources: Input Specifications, which includes the user’s chosen calibration standards;
General Specifications, which includes the user’s chosen processes; and Performance
Specifications, which includes the user’s chosen applications for the instrumentation. The
parameters and value limits listed in the following specifications indicate the product
line’s general acceptance limits and are not a report of any unit’s specific error
contribution. Any parameter exceeding the specified limits should be considered in need
of maintenance.

Specifications

Pressure Ranges

0-6000, 0-10,000 psia

Accuracy

0.01% FS

Resolution

0.001% FS or better

Control Stability

Typically 0.0075% FS

Control Response

20 – 90 seconds controlling into a 15 cu. in. volume
depending on full scale range and increment step size

Control Range

300 – 3000, 600 – 6000, 1000 – 10,000 psia

Control Volume

3 in

3

(50cc) to 35 in

3

(570cc) all ranges

Overshoot

0.01% of Full Scale

Overpressure Protection

Relief valve set at 110% FS

Connections

Supply Pressure port: 1/8

NPT

Test port: 1/8 NPT
Drive Air Supply: 1/4 NPT

Display

Color active matrix TFT display

Pressure Medium
(test and control gas)

Clean, dry, non-corrosive gas with a dew point of -40 °C
and a particulate size

<

100 microns.

Gas Consumption

Zero at commanded pressure with a leak tight system

Pressure Supply

110 to 120 psi gas pressure for controller circuit. Test gas
supply to be 110% higher than full scale pressure (may
require an external booster system).

Recommended
Recalibration Interval

1 year

Communications

Standard: RS-232 and IEEE-488

Dimensions

19" W x 24" D x 17" H
(49 x 61 x 43 cm)

Weight 120

lbs