Ansi/ieee 488.2-1987 command summary, Scpi command summary – Fluke RUSKA 7350 User Manual
Page 56

RUSKA 7350
Users Manual
5-4
ANSI/IEEE 488.2-1987 Command Summary
*CLS
Clear
Status
*ESE?
Event
Status
Enable
Query
*ESE
Event
Status
Enable
*ESR?
Event
Status
Register
*IDN?
Identification
*OPC?
Operation Complete Query (Returns 1)
*OPC
Operation
Complete
*RST
Reset
*SRE?
Service
Request
Enable
Query
*SRE
Service Request Enable
*STB?
Status
Byte
Query
*TST?
Self-Test
Query
*WAI
Wait
(No
operation)
SCPI Command Summary
The current value associated with a SCPI command may be read by appending a question
mark to the command. For example CALC:LIM:UPP? will return the current upper
pressure limit.
MEASure
[:PRESsure]?
returns current pressure reading
:POSITION?
returns pump position (%)
CALCulate
:LIMit
:LOWer
get/set low pressure limit
:SLEW
get/set slew rate limit
:UPPer
get/set high pressure limit
:VENT
get/set auto-vent limit
:TARE
:VALue
get/set tare value
:STATe ON/OFF
set tare using current pressure
CALibration
[:PRESsure]
:Data
:POINts ?
number of calibration constants
:VALue
returns cal constants label, value
:VALue
set calibration constant
:CALibration
:POINts?
number of calibration points
:VALue?
Nominal
calibration
point
:VALue
perform calibration point
:ZERO
performs
zero
calibration
:VALUE
sets zero value
:INITiate
enter zero calibration mode
:INITiate?
returns status for cal, press, temp, ref
:RUN
start
zero
calibration
:STOP
abort
zero
calibration
:POSition
:DATA
sets C0, C1 position
:DATA?
reads
C0,
C1
:VALue
sets first calibration point
:VALue 2
sets second calibration point