beautypg.com

Para 18 – HP 3325A User Manual

Page 19

background image

TB 9-6625-2049-35

19

Table 5. Dc Offset - Continued

Test instrument

DC OFFSET

Multimeter indications

(V dc)

settings

Min

Max

14.99 mV

0.01493

0.01505

-14.99 mV

-0.01505

-0.01493

4.999 mV

0.004979

0.005019

-4.999 mV

-0.005019

-0.004979

1.499 mV

0.001479

0.001519

-1.499 mV

-0.001519

-0.001479

NOTE

Perform (8) through (11) below for option 002 only.

(8) Remove 50

feedthrough termination from equipment setup.

(9) Press SIGNAL key on.
(10) Enter 20 VOLT using DATA keys. Multimeter will indicate between 19.775
and 20.225 V dc.
(11) Enter -20 VOLT using DATA keys. Multimeter will indicate between -19.775
and -20.225 V dc.

b. Adjustments. No adjustments can be made.

18. Dc Offset with Ac

a. Performance Check

(1) Connect SIGNAL to multimeter INPUT HI and LO using 50

feedthrough

termination.
(2) Press keys and enter values using DATA keys as listed in (a) through (i) below:
(a) SIGNAL off (option 002 not lit).
(b) FUNCTION sine wave.
(c) ENTRY FREQ.
(d) 20.999 999 999 MHz.
(e) ENTRY AMPTD.
(f) 1 VOLT.
(g) ENTRY DC OFFSET.
(h) 4.5 VOLT.
(i) AMPTD CAL.

(3) Multimeter will indicate between 4.35 and 4.65 V dc.

This manual is related to the following products: