C. calibration and measurement – Celestron Handheld Digital Microscope Pro (Gray) User Manual
Page 9
ENGLISH
9
+
The Celestron Micro CapturePro software includes a
measurement feature. To use this feature, you must first
perform a measurement calibration.
+
Make sure the X,Y Guide is set to “On” by selecting it in
the Options Menu.
+
Double click the saved image to open it in the Image
Preview screen.
+
Take a still image. The image will be displayed in the
column on the right side of the screen.
+
Click on the Calibration icon
+
Click and drag a line across a “known value” on the ruler image.
Try to keep the line as perpendicular to the lines on the ruler as
possible.
+
The “Adjust magnification” window will pop up as shown here.
+
In the box “Actual Dimension”, enter the “known value” from the
calibration ruler
+
Set the height of the microscope and focus on your
specimen. Then, replace your specimen with the
calibration ruler.
+
Re-adjust the focus and align the calibration ruler so that
the lines on the ruler are parallel with the vertical line on
the X,Y Guide.
C. CaliBration and MeasureMent