Cleaning/maintenance – Precision Medical PM50 EasyComp Compressor User Manual

Page 8

background image

7

9.

Begin procedure.

10. When procedure is complete, press the On/Off Switch to

the “OFF” position.

CLEANING/MAINTENANCE

WARNING

Turn the Easy Comp Compressor off and disconnect

from power source before cleaning.

To clean the filter (weekly):

1. Unplug Compressor.
2. Pry off filter end cap to check air filter for dust buildup.
3. If dust buildup occurs, remove air filter and wash in warm,

soapy water and rinse.

4. Absorb excess water from filter with a dry towel.
5. Return filter and end cap.

Cleaning for Exterior Components:
Use a clean cloth or sponge dampened with soap and water.

Wipe dry with a clean cloth or paper towel.

CAUTION

• Excessive dust buildup on filter will reduce performance of

Compressor. If this occurs, clean or replace with a new filter.

• The air filter does not filter out saline. Treatments

using nebulized saline will draw the saline mist back into

the Compressor. Saline cause rapid deterioration of the

compressor seals and will void the warranty.